NA 04.112.620SRecalls & reported problems
Synthes GmbH · 2.7MM TI VA LCP CLAVICLE PLATE SHAFT/ CS1/ LEFT/ STERILE · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886982279262), received between Sep 14, 2022 and Oct 21, 2025 — classified as 5 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K201321, K203414); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Adverse Event Without Identified Device or Use Problem2
- Device-Device Incompatibility2
- Defective Device1
- Migration1
Patient problems
- Insufficient Information3
- Failure of Implant2
- No Clinical Signs, Symptoms or Conditions2
- Nonunion/Delayed-union Bone Fracture2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryOct 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT AN UNKNOWN SURGERY WITH THE PRODUCTS IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. AFTER THE SURGERY, IT WAS CONFIRMED THAT A REVISION SURGERY WAS NEEDED DUE TO PSEUDARTHROSIS.
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 23349487 ↗InjuryOct 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT AN UNKNOWN SURGERY WITH THE PRODUCTS IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. AFTER THE SURGERY, IT WAS CONFIRMED THAT A REVISION SURGERY WAS NEEDED DUE TO PSEUDARTHROSIS.
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 23349471 ↗MalfunctionJun 27, 2025Break; Break
IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT AN UNKNOWN SURGERY FOR MIDSHAFT CLAVICLE FRACTURE WITH THE VACP CS1. ON (B)(6) 2025, THE SALES REP GOT AN INFORMATION THAT THE SCREWS BROKE AT THE REMOVAL SURGERY. WHEN THE SURGEON TURNED THE SCREWDRIVER TO REMOVE THE SCREWS, IT BROKE AT THE SCREW HEAD. USING PLIERS ETC., THEY WERE ALL REMOVED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 22345854 ↗InjuryFeb 6, 2025Break; Break
IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT A SURGERY WITH THE PLATE IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. AFTER THE SURGERY ON (B)(6) 2025, THE SALES REP WAS INFORMED THAT THE PLATE DISLOCATED. WHEN THE SALES REPRESENTATIVE WENT TO THE HOSPITAL TO CONFIRM THE X-RAY, IT WAS CONFIRMED THAT THE SCREW REMAINED IN THE PLATE AND THE DISTAL BONY FRAGMENT, WHICH WAS LUMPED WITH THE SHOULDER JOINT, WAS DISPLACED CAUDALLY BY THE WEIGHT OF THE ARM. NO SCREW FRACTURE DAMAGE WAS SEEN. THE REVISION SURGERY IS SCHEDULED ON (B)(6) 2025, TO REMOVE THE PLATE AND REPLACE WITH OTHER COMPANY¿S IMPLANTS. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 21323482 ↗MalfunctionNov 6, 2024Defective Device; Defective Device
ADDITIONAL INFORMATION RECEIVED STATES THAT THE PATIENT WAS STABLE AND NO OTHER MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20622435 ↗MalfunctionJun 11, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON MAY 15, 2024, THE PATIENT UNDERWENT THE SURGERY FOR CLAVICLE FRACTURE WITH THE VA CLAVICLE PLATE AND THE VA 2.7 LOCKING SCREW IN QUESTION. DURING SURGERY, THE SURGEON WAS CONCERNED ABOUT PLATE FLOATATION, SO HE INSERTED A VA 2.7 LOCKING SCREW WHILE HOLDING IT DOWN AND OBSERVED THE SCREW BREAKAGE UNDER THE SCREW HEAD. NOTHING REMAINS IN THE BODY HAVE BEEN REPORTED. THE DETAILS ARE UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME THIS REPORT IS FOR ONE (1) 2.7MM TI VA LCP CLAVICLE PLATE SHAFT/ CS1/ LEFT/ STERILE. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 19515595 ↗InjuryDec 1, 2022Migration; Migration
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT UNDERWENT OPEN REDUCTION/INTERNAL FIXATION SURGERY ON (B)(6) 2022 FOR A CLAVICLE TRUNK FRACTURE WITH THE PRODUCTS IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY DELAY. ON (B)(6) 2022, THE AGENCY CONTACTED AND VISITED THE SURGEON AND FOUND THAT THE PROXIMAL FIXATION HAD FAILED. 3DCT AND X-RAYS CONFIRMED THAT TWO SCREWS (ONE CORTICAL AND ONE LOCKING) IN THE PROXIMAL BONE FRAGMENT HAD BOUNCED UPWARD WITH THE PLATE AND WERE OUT OF THE BONE, WHILE THE REMAINING TWO LOCKING SCREWS REMAINED IN THE BONE WITH THE SCREW HEADS PENETRATING THE SCREW. THE POSITION OF THE IMPLANT APPEARED TO THE SURGEON TO BE GENERALLY FINE. THE IMPLANT HAD FAILED WHEN THE PATIENT SLEPT WITH THE AFFECTED SIDE DOWN AND WOKE UP IN THE MORNING. EXTRACTION WOULD BE PERFORMED ON (B)(6) 2022 AND THE FRACTURE WOULD BE REFIXED WITH A COMPETITOR PLATE. THIS REPORT INVOLVES ONE 2.7MM TI VA LCP CLAVICLE PLATE SHAFT/ CS1/ LEFT/ STERILE. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: Failure of Implant
MAUDE report 15893344 ↗MalfunctionOct 19, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT UNDERWENT AN OPEN REDUCTION/INTERNAL FIXATION OF THE CLAVICLE ON (B)(6) 2022. WHILE THE SURGEON WAS FIXING THE MEDIAL FRAGMENT WITH THE CS1 PLATE, THE SURGEON DRILLED, MEASURED, AND MANUALLY INSERTED THE APPROPRIATE LOCKING SCREWS WITH VA SLEEVE. HOWEVER, THE SCREW HAD A PIECE OF METAL ON IT AND WAS NOT TORQUED. SIMILARLY, WHEN THE SURGEON DRILLED, MEASURED, AND INSERTED THE RELEVANT PRODUCT WITH A VA SLEEVE INTO ANOTHER HOLE, IT WAS TORQUED AND USED WITHOUT PROBLEMS. THE SURGEON COMMENTED THAT IT HAPPENED BECAUSE THE SCREW WENT IN AT A STRONGER OBLIQUE ANGLE THAN THE DIRECTION IN WHICH IT WAS DRILLED. THE SURGEON CONFIRMED BY X-RAY THAT THERE WERE NO PIECES LEFT IN THE PATIENT. THERE WAS NO PROBLEM WITH BONE FIXATION. THERE WAS NO SURGICAL DELAY IN RELATION TO THIS EVENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE 2.7MM TI VA LCP CLAVICLE PLATE SHAFT/ CS1/ LEFT/ STERILE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15632015 ↗InjurySep 14, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT SURGERY TREATING THE CLAVICLE. AFTER SURGERY, ON (B)(6) 2022, IT WAS REPORTED THAT THE SCREWS BROKE. A REVISION SURGERY IS PLANNED ON (B)(6) 2022. IN THE REVISION SURGERY, BROKEN IMPLANTS WILL BE EXTRACTED, AND A LONG PLATE WILL BE USED FOR TREATMENT. THIS REPORT INVOLVES ONE 2.7MM TI VA LCP CLAVICLE PLATE SHAFT/ CS1/ LEFT/ STERILE. THIS IS REPORT 1 OF 4 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant
MAUDE report 15417143 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K201321, K203414. One submission often covers many models, so check the product and lot details in each record.
- Z-1241-2024 · Jan 24, 2024 · Open, ClassifiedVA-LCP Clavicle Plate 2.7, shaft, CS1, right-Intended for fixation of clavicle bone fragments. Part Number: 02.112.621S
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page