Skip to content

NA 04.112.621SRecalls & reported problems

Synthes GmbH · 2.7MM TI VA LCP CLAVICLE PLATE SHAFT/ CS1/ RIGHT/ STERILE · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982279293), received between Jan 9, 2023 and Jun 4, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

1 other recall cover devices cleared under the same FDA submission (K201321, K203414); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1
  • Migration1

Patient problems

  • Failure of Implant1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 4, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT ON (B)(6) 2024, THE PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION (ORIF) SURGERY FOR A FRACTURE OF THE DISTAL END OF THE CLAVICLE. ON AN UNKNOWN DATE AFTER THE SURGERY, THE SCREW BACKED OUT. IT WAS ALSO CONFIRMED THAT ONLY THREE SCREWS HAD BEEN INSERTED INTO THE PERIPHERAL BONE FRAGMENT ALTHOUGH FOUR SCREWS SHOULD HAVE BEEN INSERTED. ON (B)(6) 2024, A REVISION SURGERY WAS PERFORMED, AND IT WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. IT IS BELIEVED THAT THE IMPLANTS WERE OVERLOADED BECAUSE OF THE INSUFFICIENT NUMBER OF SCREWS AND THE FARM WORK THE PATIENT PERFORMED AFTER SURGERY. THIS REPORT IS FOR A 2.7MM TI VA LCP CLAVICLE PLATE SHAFT/ CS1/ RIGHT/ STERILE. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 19462901 ↗
  • InjuryJan 9, 2023Migration; Migration

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT ORIF FOR CLAVICULAR DIAPHYSIS WITH THE PRODUCTS IN QUESTION. TWO MONTHS LATER ON (B)(4) 2022, IT BACKED OUT. THE DOCTOR HAS CONFIRMED BY X-RAY THAT THERE ARE NO BROKEN PIECES OR OTHER RESIDUALS IN THE BODY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 2.7MM TI VA LCP CLAVICLE PLATE SHAFT/ CS1/ RIGHT/ STERILE. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: Failure of Implant

    MAUDE report 16121073 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201321, K203414. One submission often covers many models, so check the product and lot details in each record.

  • Z-1241-2024 · Jan 24, 2024 · Open, Classified
    VA-LCP Clavicle Plate 2.7, shaft, CS1, right-Intended for fixation of clavicle bone fragments. Part Number: 02.112.621S

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page