Skip to content

NA 04.045.034SRecalls & reported problems

Synthes GmbH · LOCKING SCREW FOR IM NAIL Ø 5MM/ 34MM/ XL25/ STERILE · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982294395), received between Aug 18, 2023 and Oct 23, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K201336, K201346); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem1
  • No Apparent Adverse Event1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PHYSICIAN REPORTED THAT A SHORT RFNA WHICH HE HAD IMPLANTED ON (B)(6) 2022 HAD BROKEN AT THE JUNCTION BETWEEN THE MOST DISTAL SCREW OF THE PROXIMAL INTERLOCKING SCREWS. THE SURGEON REMOVED THE BROKEN HARDWARE ON (B)(6) 2022 AND REVISED THE CONSTRUCT WITH A NAIL AND PLATE. UPON PRODUCT RETURN TO THE MANUFACTURER AND EXAMINATION ON OCTOBER 11, 2023, IT WAS IDENTIFIED THAT TWO SCREWS WERE STRIPPED. THIS REPORT INVOLVES ONE LOCKING SCREW FOR IM NAIL Ø 5MM/ 34MM/ XL25/ STERILE. THIS IS REPORT 3 OF 3 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 17984191 ↗
  • MalfunctionAug 18, 2023Tear, Rip or Hole in Device Packaging; Manufacturing, Packaging or Shipping Problem; Tear, Rip or Hole in Device Packaging; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE RECEIVED PACKAGE WAS TAPED TOGETHER, DELIVERED BY UPS. THE TOP OF THE PACKAGE WAS COMPLETELY TORN OFF AND TAPED TOGETHER. THE CONTENTS SUSTAINED PHYSICAL DAMAGE. THE AUDIT OF THE PACK SLIP SHOWED TO ONE SCREW AND NAIL MISSING. THE PACKING CONTAINER IS UNABLE TO MAINTAIN INTEGRITY DURING TRANSIT. THERE WAS NO PATIENT CONSEQUENCES/OUTCOME. THIS REPORT IS FOR ONE (1) LOCKING SCREW FOR IM NAIL Ø 5MM/ 34MM/ XL25/ STERILE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17574112 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201336, K201346. One submission often covers many models, so check the product and lot details in each record.

  • Z-0229-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A
  • Z-0228-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates