Skip to content

NA 04.043.145SRecalls & reported problems

Synthes GmbH · TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982297655), received between Dec 14, 2022 and Dec 4, 2025 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).

2 other recalls cover devices cleared under the same FDA submission (K201336); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility5
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Nonunion/Delayed-union Bone Fracture1
  • Unspecified Infection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 4, 2025Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT THE TIBIAL NAIL IMPLANT REMOVAL PROCEDURE DUE TO BREAKAGE. PER SURGEON THE BREAK WAS LIKELY DUE TO THE +3 MONTH NON-UNION OF A COMPLEX PROXIMAL TIBIAL FRACTURE. THE BROKEN NAIL WAS SUCCESSFULLY REMOVED AND A NEW TIBIAL NAIL WITH ADDITIONAL PLATING AND BONE GRAFT WAS INSERTED INTO THE PATIENT.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 23720711 ↗
  • InjurySep 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED ON (B)(6) 2025 THAT, THE PATIENT UNDERWENT HARDWARE REMOVAL TNA. ALL AVAILABLE INFORMATION WAS DISCLOSED FOR REPORTING. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Unspecified Infection

    MAUDE report 23069376 ↗
  • MalfunctionOct 30, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2024 THE REAMING GUIDE ROD KEPT TRYING TO GO POSTERIOR TO THE PROXIMAL PEEK INLAY INSTEAD OF THROUGH THE CANNULATION. AFTER MULTIPLE ATTEMPTS TO GET THE REAMING ROD TO GO THROUGH, IT WAS DECIDED TO PULL THE NAIL OUT, REMOVE THE PEEK INLAY, AND RE-IMPLANT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED AND THERE WAS 15 MINUTES OF SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20567052 ↗
  • MalfunctionJun 6, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2024 , WHILE IMPLANTING A TIBIAL NAIL ADVANCED 9 X 375MM THE 2.5MM REAMING ROD BECAME STUCK INSIDE THE NAIL WHEN TRYING TO REMOVE THE REAMING ROD. THE NAIL HAD TO BE REMOVED WITH THE REAMING ROD STILL STUCK IN IT. IT WAS REPLACED WITH A REAMING ROD AND NAIL OF THE SAME SIZES. PROLONGED ANESTHESIA, HIGHER AMOUNT OF X-RAYS, AND HIGHER SOFT TISSUE IRRITATION. THERE WAS A SURGICAL DELAY OF 35 MINUTES. THE PROCEDURE COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19475655 ↗
  • MalfunctionMay 1, 2024Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, PATIENT HAD A DISTAL TIBIA FRACTURE. AFTER REAMING TO APPROPRIATE DIAMETER, THE SURGEON INSERTED THE TIBIAL NAIL INTO THE CANAL. AFTER THE SURGEON INSERTED THE NAIL, HE DECIDED HE NEEDED TO REMOVE THE NAIL AND REAM AGAIN AFTER PLACING A BLOCKING WIRE. UPON REMOVING, IT WAS DISCOVERED THAT THE POLY INLAY FELL OUT OF THE DISTAL PORTION OF THE TIBIAL NAIL INTO THE PATIENT'S INTER MEDULLARY CANAL. THE DAMAGED POLY INLAY WAS REMOVED SUCCESSFULLY, AND WAS UNABLE TO BE RE-INSERTED INTO THE NAIL. A NEW NAIL OF THE SAME SIZE WAS OPENED, AND INSERTED INTO THE PATIENT. THERE WERE PATIENT STATUS/ OUTCOME / CONSEQUENCES REPORTED. THIS REPORT IS FOR ONE (1) TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19226768 ↗
  • MalfunctionSep 26, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2023, DURING A TIBIAL NAIL CASE, THE SURGEON CHANGED TO A SMALLER SUPRAPATELLAR SLEEVE AS THEY WERE HAVING DIFFICULTY GETTING TO THE STARTING POINT. THE SURGEON WAS EVENTUALLY ABLE TO GET TO THE STARTING POINT AND PROCEEDED WITH THE SURGERY. WHEN PUTTING IN THE PROXIMALLY LOCKING SCREWS, THE DRILL BIT IN THE S1 SCREW HOLE HIT THE NAIL. THE SURGEON WAS UNABLE TO GET THE BIT PAST THE NAIL AND SO CHOSE OTHER SCREW HOLES WHERE THE SCREWS WERE SUCCESSFULLY INSERTED. THE SCREW HOLE ALIGNMENTS WERE CHECKED PRIOR TO NAIL INSERTION AND THEY ALL LINED UP, THE SURGEON FELT THAT THE AIMING ARM WAS NOT PROPERLY LATCHED. IT WAS DISCOVERED AT THE END OF THE SURGERY THAT THE SPRING ON THE LATCH OF THE AIMING ARM WAS NOT WORKING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE PATIENT OUTCOME. THERE WAS A SURGICAL DELAY OF 45 MINUTES. NO FRAGMENTS WERE GENERATED. NO ADDITIONAL INFORMATION IS AVAILABLE TO REPORT. THIS REPORT IS FOR A TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE. THIS IS REPORT 2 OF 3 FOR PC (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17819662 ↗
  • MalfunctionDec 14, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2022 DURING A TRAUMA PROCEDURE, THE PEEK INLAY FROM THE TIBIAL NAIL ADVANCED CAME DISLODGED FROM AN UNUSED SCREW HOLE DURING NAIL INSERTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. THIS REPORT IS FOR A TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15980435 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201336. One submission often covers many models, so check the product and lot details in each record.

  • Z-0229-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A
  • Z-0228-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page