Skip to content

NA 04.045.366TSRecalls & reported problems

Synthes GmbH · LOW PROF LCKNG SCREW F/IM NAIL Ø 5.0MM / L 66MM / XL25/ STER · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982301062), received between Apr 3, 2026 and Apr 3, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K201336, K201346); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Slipped1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 3, 2026Break; Device Slipped

    IT WAS REPORTED THAT ON (B)(6) 2026 THE PATIENT UNDERWENT A REMOVAL OF A LOOSE SCREW FROM A TNA. WHEN TAKING OUT THE PROTRUDING SCREW, IT WAS DISCOVERED THAT THE SECOND INTERLOCK WAS BROKEN. THIS WAS A MONTH TO THE DAY AFTER THE SURGERY. THIS COMPLAINT IS REGARDING BOTH PROXIMAL INTERLOCKS BREAKING WITHIN 28 DAYS OF IMPLANTATION. THEY WERE 66 AND 64 MM 5.0 LOW PROFILE ANS 5.0 INTERLOCKING SCREWS. THIS REPORT CAPTURES THE SECOND REVISION SURGERY PERFORMED ON (B)(6) 2026 DUE TWO (2) LOOSE/BROKEN PROXIMAL INTERLOCKING SCREWS WHILE RELATED COMPLAINT (B)(4) CAPTURES THE FIRST REMOVAL SURGERY PERFORMED ON (B)(6) 2026 DUE TO NONUNION/MALUNION.

    Patient problems: Insufficient Information

    MAUDE report 24780215 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201336, K201346. One submission often covers many models, so check the product and lot details in each record.

  • Z-0229-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
  • Z-0228-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.