Skip to content

VOLT 02.527.316Recalls & reported problems

Synthes GmbH · 2.7MM VOLT(TM) LOCKING SCREW 16MM T8 · Class 2 · Implant

All models of VOLT (66)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982334138), received between Feb 20, 2026 and Jun 1, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility4
  • Break2
  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 1, 2026Device-Device Incompatibility

    IT WAS REPORTED MULTIPLE SCREWS WOULD NOT LOCK INTO THE PLATE. DIFFERENT LOCKING GUIDES AND TRAJECTORIES OF THE DRILL WAS TRIED. THEY ALSO TRIED A NEW PLATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A FORTY-FIVE MINUTE DELAY, BUT IT WAS NOTED THE FIXATION WAS SUBOPTIMAL, BUT NOT OTHER MEDICAL INTERVENTION WAS NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25350507 ↗
  • MalfunctionApr 16, 2026Break; Crack; Device-Device Incompatibility

    IT WAS REPORTED ON AN UNKNOWN DATE THAT SEVERAL DEVICES WERE UNABLE TO BE STERILIZED OR ASSEMBLED; DEVICES WERE CHIPPING AND CRACKING. DEVICES WERE NOT USABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24897313 ↗
  • MalfunctionApr 16, 2026Break; Crack; Device-Device Incompatibility

    IT WAS REPORTED ON AN UNKNOWN DATE THAT SEVERAL DEVICES WERE UNABLE TO BE STERILIZED OR ASSEMBLED; DEVICES WERE CHIPPING AND CRACKING. DEVICES WERE NOT USABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24897097 ↗
  • MalfunctionFeb 20, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT SURGEON WAS TREATING A DISTAL HUMERUS FRACTURE SHOWN BELOW AND A 2.7MM VOLT LOCKING SCREW, 16MM DID NOT LOCK INTO ONE OF THE TWO HEAD HOLES OF 2.7MM VOLT CONDYLAR PLATE. ACCORDING TO SURGEON PREDRILL 2.0MM HOLE WITH SPHERICAL TIP OF DRILL GUIDE IN THE POSTERIOR LATERAL COLUMN OF DISTAL HUMERUS AND 2.7MM VOLT LOCKING SCREW, 16MM DID NOT LOCK INTO ONE OF THE HEAD HOLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24402196 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page