MatrixMANDIBLE 03.505.075Recalls & reported problems
Synthes GmbH · Ø1.5MM DRILL BIT W/6MM STOP 13MM F/90° SCRWDRVR MTRXMANDBL · Class 2
All models of MatrixMANDIBLE (79)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10887587013619), received between Jun 26, 2018 and May 1, 2023 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
- Failure to Cut1
Patient problems
- Foreign Body In Patient2
- No Clinical Signs, Symptoms or Conditions1
- Not Applicable1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2023Failure to Cut; Failure to Cut
DEVICE REPORT FROM DEPUY SYNTHESE REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A SSRO FOR JAW DEFORMITY PERFORMED ON (B)(6) 2023. DRILLING WAS PERFORMED ON THE MANDIBLE; HOWEVER, IT COULD NOT BE DRILLED. ALTHOUGH THE DRILL BIT WAS CHANGED TO DIFFERENT DRILL BITS, THE SAME RESULT WAS OBTAINED. THE SHAFT WAS CHANGED, AND DRILLING WAS DONE SUCCESSFULLY. PER THE SURGEON, THE DRILL BITS STOPPED DURING DRILLING. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES DELAY. THERE WAS NO HARM TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR ONE (1) DRILL BIT Ø1.5 W/STOP L13/6 2FLUTE THIS IS REPORT 1 OF 4 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16845310 ↗MalfunctionMay 23, 2019Break; Break
CONCOMITANT DEVICE REPORTED: UNKNOWN ELAN HANDPIECE (PART #: UNKNOWN, LOT #: UNKNOWN, QUANTITY: 1).
Patient problems: Not Applicable
MAUDE report 8637509 ↗InjuryJun 26, 2018Break; Break
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2018, THE TIPS OF TWO (2) MATRIXMANDIBLE DRILL BITS BROKE WHILE DRILLING THE HOLE. THE TIPS COULD NOT BE RECOVERED AND WERE LEFT INSIDE THE PATIENT¿S BONE. THERE WAS NO PATIENT CONSEQUENCE. IT WAS UNKNOWN IF THERE WAS A SURGICAL DELAY. PROCEDURE OUTCOME WAS UNKNOWN. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 7638678 ↗InjuryJun 26, 2018Break; Break
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2018, THE TIPS OF TWO (2) MATRIXMANDIBLE DRILL BITS BROKE WHILE DRILLING THE HOLE. THE TIPS COULD NOT BE RECOVERED AND WERE LEFT INSIDE THE PATIENT¿S BONE. THERE WAS NO PATIENT CONSEQUENCE. IT WAS UNKNOWN IF THERE WAS A SURGICAL DELAY. PROCEDURE OUTCOME WAS UNKNOWN. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 7638674 ↗
Other DRIVER, WIRE, AND BONE DRILL, MANUAL devices
- NA 03.505.003Synthes GmbH · DRIVER, WIRE, AND BONE DRILL, MANUAL154 reports
- NA 5400007000Stryker Corporation · DRIVER, WIRE, AND BONE DRILL, MANUAL20 reports
- VSP System Mandible Cutting Guide3d Systems, Inc. · Driver, Wire, And Bone Drill, Manual13 reports
- NA 03.505.007Synthes GmbH · DRIVER, WIRE, AND BONE DRILL, MANUAL6 reports
- VSP System Intermediate Surgical Splint3d Systems, Inc. · Driver, Wire, And Bone Drill, Manual6 reports
- VSP System Fibula Cutting Guide with Implant Placement3d Systems, Inc. · Driver, Wire, And Bone Drill, Manual6 reports
- VSP System Fibula Cutting Guide3d Systems, Inc. · Driver, Wire, And Bone Drill, Manual5 reports
- VSP System Maxilla Cutting Guide3d Systems, Inc. · Driver, Wire, And Bone Drill, Manual4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page