NA 03.507.002SRecalls & reported problems
Synthes GmbH · 2.4MM THREADED REDUCTION TOOL SELF-DRILLING 78MM HEX-STERILE · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10887587014166), received between Feb 14, 2019 and Jul 8, 2025 — classified as 3 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Entrapment of Device2
Patient problems
- Foreign Body In Patient3
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Injury1
- No Code Available1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 8, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT ON (B)(6), 2025 THE 2.4 THREADED REDUCTION TOOL BROKE OFF IN PATIENT WHILE THE SURGEON WAS TRYING TO REDUCE ZMC FRACTURE.
Patient problems: Foreign Body In Patient
MAUDE report 22438448 ↗MalfunctionOct 23, 2024Break; Break
IT WAS REPORTED THAT PATIENT UNDERWENT ORIF FOR FACIAL FRACTURE ON (B)(6) 2024. THE THREAD REDUCTION TOOL BROKE DURING THE PROCEDURE. IT WAS DISCARDED IN THE SHARPS BIN.
Patient problems: Insufficient Information
MAUDE report 20514130 ↗InjuryJan 22, 2024Break; Break
IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2024, THERE WAS A FAILURE OF DEVICE DURING USE. THE THREADED REDUCTION TOOL BROKE OFF INTO THE PATIENT'S BONE AND REQUIRED REMOVAL. A LARGER INCISION WAS REQUIRED TO BE MADE ON THE CHEEK OF THE PATIENT IN ORDER TO REMOVE THE BROKEN SCREW PORTION OF THE DEVICE. PROCEDURE WAS COMPLETED WITH A ONE HOUR DELAY. PATIENT IS CURRENTLY IN GOOD HEALTH. THIS REPORT IS FOR A 2.4MM THREADED REDUCTION TOOL SELF-DRILLING 78MM HEX-STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 18553917 ↗MalfunctionOct 3, 2023Break; Break
IT WAS REPORTED THAT THE SURGEON REDUCED ZMC FX USING 2.4 THRD RED TOOL SELF-DRILLING 78 HEX-S. FX REDUCED AND WHILE MANIPULATING FOR FINE ADJUSTMENTS TO REDUCE THE FX A PORTION OF THE 2.4 THREADED REDUCTION TOOL BROKE OFF IN THE BONE. AFTER 40MIN MINIMUM THE END OF THE THREADED PORTION OF THE TOOL WAS REMOVED WITHOUT ANY COMPLICATION DELAYING THE CASE INTRAOPERATIVELY. THE PART HAS BEEN GIVEN. THERE WAS A PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) (B)(4). THIS IS REPORT 1 OF 1 FOR COMPLAINT 2.4 THRD RED TOOL SELF-DRILLING 78 HEX-S.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17861425 ↗MalfunctionJun 18, 2021Break; Break
IT WAS REPORTED THAT ON (B)(6) 2021, THE THREADED REDUCTION TOOL BREAK INSIDE THE PATIENT. THE BROKEN PART WAS RETRIEVED. THERE WAS A SURGICAL DELAY. PROCEDURE OUTCOME AND PATIENT STATUS ARE UNKNOWN. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12026590 ↗InjuryMar 13, 2020Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT ON (B)(6) 2020, THE 2.4MM THREADED REDUCTION TOOL SELF- DRILLING 78MM HEX -STERILE BROKE OFF IN THE PATIENT'S FACE. SWITCHED TO SELF-TAPPING CARROLL-GERARD, LEFT BROKEN PIECE IN THE PATIENT. THE SCREW WAS NOT REMOVED. NO FRAGMENTS GENERATED. THERE WERE 30 MINUTES OF SURGICAL DELAY. THE PROCEDURE SUCCESSFULLY COMPLETED. THE PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNK - PLATES: MATRIXMID FACE (PART # UNKNOWN; LOT # UNKNOWN; QUANTITY: UNKNOWN). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS IS 1 OF 1 FOR REPORT (B)(4).
Patient problems: Foreign Body In Patient; Injury
MAUDE report 9832209 ↗InjuryFeb 14, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, DURING A ZYGOMATIC FRACTURE, THE THREADED REDUCTION TOOL SELF-DRILLING BROKE OFF AS IT WAS ALREADY SCREWED INTO THE ZYGOMA. THE SURGEON HAD PREVIOUSLY USED IT TO MANIPULATE THE BONE, BUT IT WAS NOT BEING TOUCHED AT THE TIME IT BROKE. THE SURGEON WAS LOOKING AT PLATES AND THE TOOL BROKE OFF AND LANDED ON THE GROUND. PART OF THE SCREW WAS UNABLE TO BE ACCESSED FOR REMOVAL FROM THE PATIENT. FRAGMENTS WERE GENERATED AND WERE NOT ACCESSIBLE FOR REMOVAL. THERE WAS NO SURGICAL DELAY AND PATIENT CONSEQUENCE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient; No Code Available
MAUDE report 8341371 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.