NA 04.500.200Recalls & reported problems
Synthes GmbH · 1.3MM CURVILINEAR DISTRACTOR R100 / RIGHT · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10887587016818), received between Jul 17, 2018 and Oct 24, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Difficult to Open or Close” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Open or Close1
Patient problems
- Bone Fracture(s)1
- Nerve Damage1
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT BILATERAL MANDIBULAR OSTEOTOMIES, APPLICATION OF BILATERAL INTERNAL DISTRACTION DEVICES, DENTAL EXTRACTIONS OF #2, 20, 31, AND A, AND RIGHT MANDIBULAR ANTROSTOMY, CAUDWELL-LUC APPROACH, TO RETRIEVE DENTAL ROOT OF #2. INTRAOPERATIVE COMPLICATIONS INCLUDED DENTAL ROOT FRACTURE REQUIRING MAXILLARY ANTROSTOMY ON THE RIGHT FOR RETRIEVAL, AND GREENSTICK FRACTURE WITH REMOVAL. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE 1.3MM CURVILINEAR DISTRACTOR R100 / RIGHT. THIS IS REPORT 1 OF 12 FOR (B)(4).
Patient problems: Bone Fracture(s); Nerve Damage
MAUDE report 15658330 ↗InjuryJul 17, 2018Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN OSTEOTOMY PROCEDURE ON (B)(46 2018. THE SURGEON PLACED A CURVILINEAR DISTRACTOR ON THE MANDIBLE AND FIXATED IT WITH SCREWS. DISTRACTOR TESTING WAS PERFORMED TO CONFIRM OSTEOTOMY COMPLETION. THE DISTRACTOR COULD ACTIVATE FORWARD BUT COULD NOT REVERSE AS THE DISTRACTOR WAS DESIGNED TO DO. THE SURGEON TRIED TO DETERMINE IF BONE OR SOMETHING ELSE WAS BLOCKING THE DISTRACTOR FROM REVERSING BUT THERE WAS NO OBSTACLE. THE SURGEON DECIDED TO REMOVE THE DISTRACTOR. AFTER REMOVING THE DISTRACTOR, THE SURGEON IRRIGATED THE DISTRACTOR HEAVILY TO CONFIRM THERE WAS NOTHING CAUSING A BLOCKAGE. THE SURGEON THEN ACTIVATED THE DISTRACTOR FORWARD AND IT PERFORMED AS NORMAL, BUT THE DISTRACTOR COULD STILL NOT REVERSE. THE SURGEON THEN DETERMINED THE DISTRACTOR WAS BROKEN AND CHOSE A SMALLER DISTRACTOR (70MM) FROM THE SAME SET. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF 1 HOUR DUE TO THE REPORTED EVENT. PATIENT STATUS IS STABLE. THIS REPORT IS FOR A 1.3MM CURVILINEAR DISTRACTOR R100 / RIGHT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7695903 ↗
FDA clearance and classification
Cleared through 510(k) K121502 on Aug 23, 2012 (decision: Substantially Equivalent). Submitted by Synthes, Inc. as “Synthes Curvilinear Distraction System”. The FDA review took 94 days. FDA record ↗
The FDA regulates product code MQN (external mandibular fixator and/or distractor) under 21 CFR 872.4760 as a Class II device. 21 CFR 872.4760 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 04.500.200
Has the NA 04.500.200 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10887587016818), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 04.500.200?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jul 17, 2018 and Oct 24, 2022. In those reports, the most common specific device problem is “Difficult to Open or Close” (1); the most common patient problems are “Bone Fracture(s)” (1) and “Nerve Damage” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 04.500.200?
None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 04.500.200 FDA approved?
It was cleared through 510(k) K121502 on Aug 23, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the NA 04.500.200 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the NA 04.500.200, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 04.500.200). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices
- DISTRACTION, INTERNAL 51-405-42-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR20 reports
- LEVEL ONE CMF 50-760-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR18 reports
- LEVEL ONE CMF 50-797-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- DISTRACTION, INTERNAL 51-563-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-777-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR11 reports
- NA 04.315.125Synthes GmbH · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-526-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-401-91-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page