NA 04.500.241Recalls & reported problems
Synthes GmbH · 1.3MM CURVILINEAR DISTRACTOR R40/LEFT · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10887587016870), received between Jan 30, 2017 and Aug 4, 2017 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected2
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 4, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED PATIENT WAS IMPLANTED WITH A 1.3MM CURVILINEAR DISTRACTOR ON (B)(6)2017. SURGEON STATES IT WAS NECESSARY TO OVER-ADVANCE THE DISTRACTOR IN ORDER TO OBTAIN FULL DISTRACTION DISTANCE. DEVICE WAS UTILIZED TO COMPLETE THE TREATMENT PLAN AND WAS REMOVED AS PLANNED ON (B)(6) 2017. SURGEON STATED THE DISTRACTOR TAB THAT PREVENTS REVERSING APPEARS TO BE DEFLECTED OUT OF THE WAY PREVENTING IT FROM ENGAGING. CONCOMITANT DEVICES REPORTED: ACTIVATION INSTRUMENT FOR CURVILINEAR DISTRACTOR (PART 03.500.016, LOT NUMBER UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) 1.3MM CURVILINEAR DISTRACTOR/LEFT. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6767347 ↗MalfunctionJan 30, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT TWO (2) 1.3MM CURVILINEAR DISTRACTORS RIGHT SIDE AND ONE (1) 1.3MM CURVILINEAR DISTRACTOR LEFT SIDE MALFUNCTIONED INTRAOPERATIVELY CAUSING A 25 MINUTE SURGICAL DELAY. A PATIENT WITH A TRACHEOSTOMY UNDERWENT A SURGICAL PROCEDURE ON (B)(6) 2017 TO TREAT A BLOCKED AIRWAY. DURING TESTING OF THE FIRST RIGHT AND LEFT SIDE DISTRACTOR DEVICES PRIOR TO INSERTION INTO THE PATIENT, THE TRACKS WERE NOT PROPERLY BACKING UP (CLOSING) AFTER BEING ADVANCED. THERE WAS AN ENORMOUS AMOUNT OF RESISTANCE THAT WAS NOT NORMAL. THESE TWO (2) DEVICES WERE NOT IMPLANTED IN THE PATIENT. ANOTHER RIGHT DISTRACTOR WAS IMPLANTED. ADVANCING FORWARD WAS FINE BUT THERE WAS DIFFICULTY GOING BACKWARDS. A LEFT DISTRACTOR WAS THEREFORE INSERTED INTO THE PATIENT. THERE WAS NO ALLEGATION OF COMPLAINT AGAINST THE SECOND LEFT DISTRACTOR. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT OUTCOME WAS STABLE. THIS REPORT IS 3 OF 3 FOR (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6286465 ↗
Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices
- DISTRACTION, INTERNAL 51-405-42-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR20 reports
- LEVEL ONE CMF 50-760-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR18 reports
- DISTRACTION, INTERNAL 51-563-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-797-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-777-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR11 reports
- DISTRACTION, INTERNAL 51-526-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- NA 04.315.125Synthes GmbH · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-401-91-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page