NA 04.500.251Recalls & reported problems
Synthes GmbH · 1.3MM CURVILINEAR DISTRACTOR R50/LEFT · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10887587016894), received between Jan 2, 2019 and Feb 10, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Mechanical Jam” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Mechanical Jam1
- Mechanical Problem1
Patient problems
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2020Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2019, THE IMPLANT DISTRACTION DEVICE DID NOT REALLY WORK LIKE IT WAS SUPPOSED TO. THERE WAS SOME PATIENT ERROR. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PATIENT STATUS AND THE OUTCOME ARE UNKNOWN. THIS IS REPORT 01 OF 04 OF (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9687008 ↗InjuryJan 2, 2019Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT A REVISION OF MANDIBULAR DISTRACTION BECAUSE THE CURVILINEAR DISTRACTOR STOPPED DISTRACTING SOMEWHERE BETWEEN 20-25MM. THE RIGHT ADVANCED FARTHER, DESPITE A SLOWER RATE OF DISTRACTION, AS COMPARED TO THE LEFT. THE REMOVABLE EXTENSION ARM-FLEXIBLE WAS IDENTIFIED TO BE STUCK AND COULD NOT BE DISASSEMBLED FROM THE CURVILINEAR DISTRACTOR. THE SURGEON REPLACED THE OLD DISTRACTOR, WITH A NEW ONE, TO FINISH THE DISTRACTION ON THE MANDIBLE. ORIGINAL IMPLANT DATE IS ON (B)(6) 2018. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PROCEDURE OUTCOME AND PATIENT STATUS ARE UNKNOWN. THIS REPORT IS FOR ONE (1) 1.3MM CURVILINEAR DISTRACTOR R50/LEFT. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 8210051 ↗
Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices
- DISTRACTION, INTERNAL 51-405-42-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR20 reports
- LEVEL ONE CMF 50-760-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR18 reports
- LEVEL ONE CMF 50-797-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- DISTRACTION, INTERNAL 51-563-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-777-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR11 reports
- DISTRACTION, INTERNAL 51-401-91-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-526-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- NA 04.315.125Synthes GmbH · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.