Skip to content

MatrixRIB 04.501.018.01Recalls & reported problems

Synthes GmbH · 2.9MM TI MATRIXRIB LCKNG SCREW SELF-TAPPING/8MM · Class 2 · Implant

All models of MatrixRIB (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10887587017150), received between Oct 10, 2022 and May 8, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility3
  • Break1
  • Device Slipped1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Break

    IT WAS REPORTED THAT THIS EVENT INVOLVED 1 SCREW AND 1 PLATE. IT WAS REPORTED THAT DURING THE SURGERY, METAL DEBRIS APPEARS DURING TIGHTENING THE SCREW TO THE LAST TURN. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25122228 ↗
  • InjuryAug 11, 2025Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION RECEIVED FROM SALES REP VIA PHONE TODAY STATES THAT AFTER INVESTIGATION, THE PATIENT UNDERWENT RIB FRACTURE INTERNAL FIXATION ON (B)(6) 2024 DUE TO "RIB FRACTURE", AND 04.501.018.01 AND 04.501.096 WERE IMPLANTED DURING THE SURGERY. THE SURGICAL PROCESS AND POSTOPERATIVE REHABILITATION OF THE PATIENT WERE UNKNOWN. ON (B)(6) 2025, THE PATIENT UNDERWENT RIB FRACTURE INTERNAL FIXATION REMOVAL. DURING REMOVING THE SCREWS AND PLATES WITH MATCHING TOOLS, THE TWO SCREWS WERE RESPECTIVELY EMBEDDED ON THE TWO PLATES AND COULD NOT BE SCREWED OUT. FINALLY, THE SURGEON HAVE TO PULL OUT THE PLATE AND SCREWS AT THE SAME TIME. THE SURGICAL TIME WAS EXTENDED DURING THIS PERIOD BY APPROXIMATELY 1.5 HOURS. THIS EVENT DID NOT CAUSE ANY OTHER HARM TO THE PATIENT EXCEPT FOR THE PROLONGED OPERATION TIME. THE PATIENT HAS BEEN DISCHARGED SMOOTHLY CURRENTLY. NO SUBSEQUENT ADVERSE EVENT REPORT WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22771548 ↗
  • InjuryAug 11, 2025Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION RECEIVED FROM SALES REP VIA PHONE TODAY STATES THAT AFTER INVESTIGATION, THE PATIENT UNDERWENT RIB FRACTURE INTERNAL FIXATION ON (B)(6) 2024 DUE TO "RIB FRACTURE", AND 04.501.018.01 AND 04.501.096 WERE IMPLANTED DURING THE SURGERY. THE SURGICAL PROCESS AND POSTOPERATIVE REHABILITATION OF THE PATIENT WERE UNKNOWN. ON (B)(6) 2025, THE PATIENT UNDERWENT RIB FRACTURE INTERNAL FIXATION REMOVAL. DURING REMOVING THE SCREWS AND PLATES WITH MATCHING TOOLS, THE TWO SCREWS WERE RESPECTIVELY EMBEDDED ON THE TWO PLATES AND COULD NOT BE SCREWED OUT. FINALLY, THE SURGEON HAVE TO PULL OUT THE PLATE AND SCREWS AT THE SAME TIME. THE SURGICAL TIME WAS EXTENDED DURING THIS PERIOD BY APPROXIMATELY 1.5 HOURS. THIS EVENT DID NOT CAUSE ANY OTHER HARM TO THE PATIENT EXCEPT FOR THE PROLONGED OPERATION TIME. THE PATIENT HAS BEEN DISCHARGED SMOOTHLY CURRENTLY. NO SUBSEQUENT ADVERSE EVENT REPORT WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22771423 ↗
  • MalfunctionNov 7, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS AN UNKNOWN SURGERY PERFORMED ON OCTOBER 19, 2023. DURING SURGERY, THE PLATE AND SCREW GOT STUCK, AND THEY BECAME UNABLE TO BE USED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. PATIENT STATUS/ OUTCOME: STABLE. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1)LOCK SCREW ø2.9 SELF-TAP L8 TAN. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18083245 ↗
  • MalfunctionOct 10, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, A SCREW WAS FOUND TO BE LOOSE DURING A POST-OP EXAMINATION. THE INITIAL SURGERY FOR TREATING THE RIB WAS PERFORMED ON (B)(6) 2022. THE PATIENT HAS NO PAIN OR IRRITATION/UNCOMFORTABLE FEELING. AT PRESENT, IT IS UNLIKELY THAT THE SCREW WOULD FALL INTO THE THORACIC CAVITY, AND THE PATIENT DID NOT DEVELOP ANY SUBJECTIVE SYMPTOMS, SO IT WAS DECIDED THAT THE PATIENT WOULD BE FOLLOWED UP AT THE DISCRETION OF THE SURGEON THIS REPORT INVOLVES ONE (1) 2.9MM TI MATRIXRIB LCKNG SCREW SELF-TAPPING/8MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15569851 ↗

Other PLATE, FIXATION, BONE devices