NA 391.990Recalls & reported problems
Synthes GmbH · PLATE/ROD CUTTER · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10887587041087), received between Mar 4, 2019 and Sep 17, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Difficult to Open or Close1
- Material Deformation1
- Material Integrity Problem1
- Mechanical Problem1
Patient problems
- No Patient Involvement2
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2024Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE ARTICLE DOES NOT CLOSE PROPERLY, SO NO THREADS CAN BE PINCHED OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20236496 ↗MalfunctionJul 27, 2020Difficult to Open or Close; Material Deformation; Difficult to Open or Close; Material Deformation
VETERINARY COMPLAINT: IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING AN UNKNOWN CASE IT WAS DISCOVERED THAT THE PLATE ROD CUTTER ARE NOT SPRINGING BACK OPEN AFTER USE. THESE CUTTERS PRIMARILY USED WITH TIBIAL PLATEAU LEVELING OSTEOTOMY (TPLO) AND TO CUT THE THREADED PIN FOR STANDARD JIG. NO FURTHER INFORMATION PROVIDED. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT THE CUTTING EDGES WERE DULL AND SLIGHTLY DEFORMED ON BOTH JAWS. THIS DEVICE CONDITION WAS REASSESSED AND DETERMINED AS REPORTABLE ON (B)(6) 2020. CONCOMITANT DEVICES REPORTED: KIRSCHNER WIRE (PART # VW3005.15, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR 1 PLATE/ROD CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10329360 ↗MalfunctionApr 20, 2020Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PLATE/ROD CUTTER DID NOT FUNCTION DURING REVERSE LOGISTICS AUDIT OF RETURNED DEVICE AT MILLSTONE. THERE WAS NO PATIENT INVOLVEMENT AND NO ADDITIONAL INFORMATION IS AVAILABLE. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE, IT WAS IDENTIFIED THAT ONE OF THE THREE PIVOT SCREWS IS MISSING ON THE PLATE/ROD CUTTER. THIS PRODUCT CONDITION WAS REASSESSED AND DETERMINED AS REPORTABLE ON MARCH 27, 2020. THIS REPORT INVOLVES ONE (1) PLATE/ROD CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 9979005 ↗MalfunctionMar 4, 2019Break; Break
IT WAS REPORTED THAT ON FEBRUARY 8, 2019, DURING TESTING AT SERVICE AND REPAIR, THE CUTTER WAS BROKEN AND DID NOT CUT. THE ISSUE WAS FOUND WHILE RECEIVING LOANER SET IN LOANER DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) PLATE/ROD CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8386201 ↗
Other INSTRUMENT, CUTTING, ORTHOPEDIC devices
- SIGMA LCS 9505-01-230Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC95 reports
- ATTUNE 2544-00-522Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC58 reports
- SIGMA LCS 9505-01-229Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC56 reports
- SIGMA 2024-40-100Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC43 reports
- ATTUNE 2545-00-158Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC41 reports
- ATTUNE 2545-00-156Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC41 reports
- SIGMA 2024-40-200Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC40 reports
- ATTUNE 2545-00-155Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC39 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page