NA 460.046Recalls & reported problems
Synthes GmbH · TI STERNAL LOCKING STRAIGHT PLATE/13 HOLES/WITHOUT PIN · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10887587055251), received between Dec 26, 2018 and Jan 30, 2020 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break4
Patient problems
- Bone Fracture(s)2
- Impaired Healing2
- Inflammation1
- No Consequences Or Impact To Patient1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 30, 2020Break; Break
IT WAS REPORTED ON (B)(6) 2020, ONE (1) STEPPED DRILL BIT CANNULATED DID NOT FUNCTION; SIX (6) DEPTH GAUGE FOR LOCKING SCREWS, TWO (2) LOCKING BOLT MEASURING DEVICE, AND ONE (1) TITANIUM STERNAL LOCKING STRAIGHT PLATE WERE FOUND TO BE BROKEN DURING REVERSE LOGISTICS AUDIT OF RETURNED DEVICE AT MILLSTONE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR (1) TI STERNAL LOCKING STRAIGHT PLATE/13 HOLES/WITHOUT PIN. THIS IS REPORT 07 OF 10 OF (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9647726 ↗InjuryJan 15, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT A HARDWARE REMOVAL PROCEDURE OF TWO (2) TITANIUM STERNAL LOCKING STRAIGHT PLATES AND TWENTY-THREE (23) UNKNOWN SCREWS DUE TO NONUNION AND BROKEN PLATES. IT WAS NOTED THAT THE PLATES AND SCREWS CAME OUT OF THE PATIENT. INITIALLY, THE PATIENT HAD AN OPEN REDUCTION INTERNAL FIXATION PROCEDURE (ORIF) WITH IMPLANTATION OF THE PLATES AND UNKNOWN SCREWS ON (B)(6) 2015, TO REPAIR AN EXTERNAL FRACTURE. IT IS UNKNOWN IF THERE WERE FRAGMENTS FROM THE BROKEN DEVICES. PATIENT AND PROCEDURE OUTCOME ARE UNKNOWN. CONCOMITANT REPORTED DEVICES: UNKNOWN SCREWS (PART # UNKNOWN, LOT NUMBER#, QUANTITY: 23). THIS REPORT IS FOR ONE (1) TI STERNAL LOCKING STRAIGHT PLATE/13 HOLES/WITHOUT PIN. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Bone Fracture(s); Impaired Healing
MAUDE report 8249396 ↗InjuryJan 15, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT A HARDWARE REMOVAL PROCEDURE OF TWO (2) TITANIUM STERNAL LOCKING STRAIGHT PLATES AND TWENTY-THREE (23) UNKNOWN SCREWS DUE TO NONUNION AND BROKEN PLATES. IT WAS NOTED THAT THE PLATES AND SCREWS CAME OUT OF THE PATIENT. INITIALLY, THE PATIENT HAD AN OPEN REDUCTION INTERNAL FIXATION PROCEDURE (ORIF) WITH IMPLANTATION OF THE PLATES AND UNKNOWN SCREWS ON (B)(6) 2015, TO REPAIR AN EXTERNAL FRACTURE. IT IS UNKNOWN IF THERE WERE FRAGMENTS FROM THE BROKEN DEVICES. PATIENT AND PROCEDURE OUTCOME ARE UNKNOWN. CONCOMITANT REPORTED DEVICES: UNKNOWN SCREWS (PART # UNKNOWN, LOT NUMBER#, QUANTITY: 23). THIS REPORT IS FOR ONE (1) TI STERNAL LOCKING STRAIGHT PLATE/13 HOLES/WITHOUT PIN. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Bone Fracture(s); Impaired Healing
MAUDE report 8249394 ↗InjuryDec 26, 2018Break; Break
CONCOMITANT DEVICES: TI STERNAL LOCKING STRAIGHT PLATE/13 HOLES/WITHOUT PIN (PRODUCT CODE: 460.046, LOT NUMBER: 7260429, QUANTITY: 1). UNK - SCREWS: TRAUMA (PRODUCT CODE: UNKNOWN, LOT NUMBER: UKNOWN, QUANTITY: 13).
Patient problems: Inflammation; Pain
MAUDE report 8197930 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page