Skip to content

MatrixRIB 04.501.213.01Recalls & reported problems

Synthes GmbH · 2.7MM MATRIXRIB TI LCKNG SCREW SELF-DRILLING/13MM · Class 2 · Implant

All models of MatrixRIB (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10887587074351), received between Feb 7, 2023 and Jun 16, 2025 — classified as 9 injuries.

The most frequently reported specific device problem is “Migration” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 8 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration8
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Pain8
  • Unspecified Infection1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 16, 2025Migration; Migration

    IT WAS REPORTED REVISION SURGERY OCCURRED ON (B)(6) 2025 DUE TO DEVICE MIGRATION AND PATIENT REPORT OF PAIN. THE SURGEON ORIGINALLY PLATED PATIENT'S 7TH RIB ON (B)(6) 2024 WITH STERNAL PLATE FOR SURGICAL STABILIZATION OF RIB FRACTURE. DURING THE REVISION, THE SURGEON HAD TO REMOVE THE PLATE AS THE PLATE AND SCREWS WERE BACKING OUT OF THE BONE AND COMING AWAY FROM THE RIB. MEDIAL SCREWS WERE REMOVED WITH A DRIVER WHILE THE LATERAL SCREWS JUST PULLED OUT OF THE BONE. THE SURGEON BELIEVES IT IS DUE TO UNDER CONTOURING OF THE PLATE. IT WAS NOTED THE SURGEON WAS FOLLOWING THE CORRECT SURGICAL TECHNIQUE IN THE ORIGINAL SURGERY. IT WAS FURTHER MENTIONED IT WAS NOT CLEAR IF THE PATIENT HAD BEEN COMPLIANT POST INITIAL PROCEDURE. THERE WAS CALLUS NOTED IN THE BONE DURING THE REVISION SO NO FURTHER DEVICES WERE IMPLANTED.

    Patient problems: Pain

    MAUDE report 22212860 ↗
  • InjuryJun 16, 2025Migration; Migration

    IT WAS REPORTED REVISION SURGERY OCCURRED ON (B)(6) 2025 DUE TO DEVICE MIGRATION AND PATIENT REPORT OF PAIN. THE SURGEON ORIGINALLY PLATED PATIENT'S 7TH RIB ON (B)(6) 2024 WITH STERNAL PLATE FOR SURGICAL STABILIZATION OF RIB FRACTURE. DURING THE REVISION, THE SURGEON HAD TO REMOVE THE PLATE AS THE PLATE AND SCREWS WERE BACKING OUT OF THE BONE AND COMING AWAY FROM THE RIB. MEDIAL SCREWS WERE REMOVED WITH A DRIVER WHILE THE LATERAL SCREWS JUST PULLED OUT OF THE BONE. THE SURGEON BELIEVES IT IS DUE TO UNDER CONTOURING OF THE PLATE. IT WAS NOTED THE SURGEON WAS FOLLOWING THE CORRECT SURGICAL TECHNIQUE IN THE ORIGINAL SURGERY. IT WAS FURTHER MENTIONED IT WAS NOT CLEAR IF THE PATIENT HAD BEEN COMPLIANT POST INITIAL PROCEDURE. THERE WAS CALLUS NOTED IN THE BONE DURING THE REVISION SO NO FURTHER DEVICES WERE IMPLANTED.

    Patient problems: Pain

    MAUDE report 22212704 ↗
  • InjuryJun 16, 2025Migration; Migration

    IT WAS REPORTED REVISION SURGERY OCCURRED ON (B)(6) 2025 DUE TO DEVICE MIGRATION AND PATIENT REPORT OF PAIN. THE SURGEON ORIGINALLY PLATED PATIENT'S 7TH RIB ON (B)(6) 2024 WITH STERNAL PLATE FOR SURGICAL STABILIZATION OF RIB FRACTURE. DURING THE REVISION, THE SURGEON HAD TO REMOVE THE PLATE AS THE PLATE AND SCREWS WERE BACKING OUT OF THE BONE AND COMING AWAY FROM THE RIB. MEDIAL SCREWS WERE REMOVED WITH A DRIVER WHILE THE LATERAL SCREWS JUST PULLED OUT OF THE BONE. THE SURGEON BELIEVES IT IS DUE TO UNDER CONTOURING OF THE PLATE. IT WAS NOTED THE SURGEON WAS FOLLOWING THE CORRECT SURGICAL TECHNIQUE IN THE ORIGINAL SURGERY. IT WAS FURTHER MENTIONED IT WAS NOT CLEAR IF THE PATIENT HAD BEEN COMPLIANT POST INITIAL PROCEDURE. THERE WAS CALLUS NOTED IN THE BONE DURING THE REVISION SO NO FURTHER DEVICES WERE IMPLANTED.

    Patient problems: Pain

    MAUDE report 22212670 ↗
  • InjuryJun 16, 2025Migration; Migration

    IT WAS REPORTED REVISION SURGERY OCCURRED ON (B)(6) 2025 DUE TO DEVICE MIGRATION AND PATIENT REPORT OF PAIN. THE SURGEON ORIGINALLY PLATED PATIENT'S 7TH RIB ON (B)(6) 2024 WITH STERNAL PLATE FOR SURGICAL STABILIZATION OF RIB FRACTURE. DURING THE REVISION, THE SURGEON HAD TO REMOVE THE PLATE AS THE PLATE AND SCREWS WERE BACKING OUT OF THE BONE AND COMING AWAY FROM THE RIB. MEDIAL SCREWS WERE REMOVED WITH A DRIVER WHILE THE LATERAL SCREWS JUST PULLED OUT OF THE BONE. THE SURGEON BELIEVES IT IS DUE TO UNDER CONTOURING OF THE PLATE. IT WAS NOTED THE SURGEON WAS FOLLOWING THE CORRECT SURGICAL TECHNIQUE IN THE ORIGINAL SURGERY. IT WAS FURTHER MENTIONED IT WAS NOT CLEAR IF THE PATIENT HAD BEEN COMPLIANT POST INITIAL PROCEDURE. THERE WAS CALLUS NOTED IN THE BONE DURING THE REVISION SO NO FURTHER DEVICES WERE IMPLANTED.

    Patient problems: Pain

    MAUDE report 22212626 ↗
  • InjuryJun 16, 2025Migration; Migration

    IT WAS REPORTED REVISION SURGERY OCCURRED ON (B)(6) 2025 DUE TO DEVICE MIGRATION AND PATIENT REPORT OF PAIN. THE SURGEON ORIGINALLY PLATED PATIENT'S 7TH RIB ON (B)(6) 2024 WITH STERNAL PLATE FOR SURGICAL STABILIZATION OF RIB FRACTURE. DURING THE REVISION, THE SURGEON HAD TO REMOVE THE PLATE AS THE PLATE AND SCREWS WERE BACKING OUT OF THE BONE AND COMING AWAY FROM THE RIB. MEDIAL SCREWS WERE REMOVED WITH A DRIVER WHILE THE LATERAL SCREWS JUST PULLED OUT OF THE BONE. THE SURGEON BELIEVES IT IS DUE TO UNDER CONTOURING OF THE PLATE. IT WAS NOTED THE SURGEON WAS FOLLOWING THE CORRECT SURGICAL TECHNIQUE IN THE ORIGINAL SURGERY. IT WAS FURTHER MENTIONED IT WAS NOT CLEAR IF THE PATIENT HAD BEEN COMPLIANT POST INITIAL PROCEDURE. THERE WAS CALLUS NOTED IN THE BONE DURING THE REVISION SO NO FURTHER DEVICES WERE IMPLANTED.

    Patient problems: Pain

    MAUDE report 22212304 ↗
  • InjuryJun 16, 2025Migration; Migration

    IT WAS REPORTED REVISION SURGERY OCCURRED ON (B)(6) 2025 DUE TO DEVICE MIGRATION AND PATIENT REPORT OF PAIN. THE SURGEON ORIGINALLY PLATED PATIENT'S 7TH RIB ON (B)(6) 2024 WITH STERNAL PLATE FOR SURGICAL STABILIZATION OF RIB FRACTURE. DURING THE REVISION, THE SURGEON HAD TO REMOVE THE PLATE AS THE PLATE AND SCREWS WERE BACKING OUT OF THE BONE AND COMING AWAY FROM THE RIB. MEDIAL SCREWS WERE REMOVED WITH A DRIVER WHILE THE LATERAL SCREWS JUST PULLED OUT OF THE BONE. THE SURGEON BELIEVES IT IS DUE TO UNDER CONTOURING OF THE PLATE. IT WAS NOTED THE SURGEON WAS FOLLOWING THE CORRECT SURGICAL TECHNIQUE IN THE ORIGINAL SURGERY. IT WAS FURTHER MENTIONED IT WAS NOT CLEAR IF THE PATIENT HAD BEEN COMPLIANT POST INITIAL PROCEDURE. THERE WAS CALLUS NOTED IN THE BONE DURING THE REVISION SO NO FURTHER DEVICES WERE IMPLANTED.

    Patient problems: Pain

    MAUDE report 22211985 ↗
  • InjuryJun 16, 2025Migration; Migration

    IT WAS REPORTED REVISION SURGERY OCCURRED ON (B)(6) 2025 DUE TO DEVICE MIGRATION AND PATIENT REPORT OF PAIN. THE SURGEON ORIGINALLY PLATED PATIENT'S 7TH RIB ON (B)(6) 2024 WITH STERNAL PLATE FOR SURGICAL STABILIZATION OF RIB FRACTURE. DURING THE REVISION, THE SURGEON HAD TO REMOVE THE PLATE AS THE PLATE AND SCREWS WERE BACKING OUT OF THE BONE AND COMING AWAY FROM THE RIB. MEDIAL SCREWS WERE REMOVED WITH A DRIVER WHILE THE LATERAL SCREWS JUST PULLED OUT OF THE BONE. THE SURGEON BELIEVES IT IS DUE TO UNDER CONTOURING OF THE PLATE. IT WAS NOTED THE SURGEON WAS FOLLOWING THE CORRECT SURGICAL TECHNIQUE IN THE ORIGINAL SURGERY. IT WAS FURTHER MENTIONED IT WAS NOT CLEAR IF THE PATIENT HAD BEEN COMPLIANT POST INITIAL PROCEDURE. THERE WAS CALLUS NOTED IN THE BONE DURING THE REVISION SO NO FURTHER DEVICES WERE IMPLANTED.

    Patient problems: Pain

    MAUDE report 22211777 ↗
  • InjuryJun 16, 2025Migration; Migration

    IT WAS REPORTED REVISION SURGERY OCCURRED ON (B)(6) 2025 DUE TO DEVICE MIGRATION AND PATIENT REPORT OF PAIN. THE SURGEON ORIGINALLY PLATED PATIENT'S 7TH RIB ON (B)(6) 2024 WITH STERNAL PLATE FOR SURGICAL STABILIZATION OF RIB FRACTURE. DURING THE REVISION, THE SURGEON HAD TO REMOVE THE PLATE AS THE PLATE AND SCREWS WERE BACKING OUT OF THE BONE AND COMING AWAY FROM THE RIB. MEDIAL SCREWS WERE REMOVED WITH A DRIVER WHILE THE LATERAL SCREWS JUST PULLED OUT OF THE BONE. THE SURGEON BELIEVES IT IS DUE TO UNDER CONTOURING OF THE PLATE. IT WAS NOTED THE SURGEON WAS FOLLOWING THE CORRECT SURGICAL TECHNIQUE IN THE ORIGINAL SURGERY. IT WAS FURTHER MENTIONED IT WAS NOT CLEAR IF THE PATIENT HAD BEEN COMPLIANT POST INITIAL PROCEDURE. THERE WAS CALLUS NOTED IN THE BONE DURING THE REVISION SO NO FURTHER DEVICES WERE IMPLANTED.

    Patient problems: Pain

    MAUDE report 22211669 ↗
  • InjuryFeb 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2023, A RIB PLATE WAS EXPLANTED. THE PLATES WERE AROUND THE PATIENT, AND WERE INFECTED SO THEY HAD TO GET REMOVED. PLATES WERE ORIGINALLY IMPLANTED ON (B)(6) 2022. THIS REPORT IS FOR A 2.7MM MATRIXRIB TI LCKNG SCREW SELF-DRILLING/13MM. THIS IS REPORT 5 OF 6 FOR (B)(4).

    Patient problems: Unspecified Infection

    MAUDE report 16311612 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page