PlumRecalls & reported problems across all models
ICU Medical, Inc. · 26 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one Plum model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1,358
- Deaths reported
- 2
- Injuries reported
- 11
- Recalls
- 6
We found 26 models of Plum by ICU Medical, Inc. in the FDA's device registry (GUDID). Together they are named in 1,358 adverse event reports in the FDA's MAUDE database, received between Aug 29, 2017 and Sep 30, 2026 — classified as 2 deaths, 11 injuries, 1,345 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (875 reports).
6 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Plum recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 12, 2026Open, Classified
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
Z-3162-2026 · Root cause: Process control · Recall details · FDA record ↗
- May 12, 2026Open, Classified
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
Z-3156-2026 · Root cause: Process control · names 9 of these models · Recall details · FDA record ↗
- Jul 29, 2019Terminated
Potential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.
Z-0145-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 29, 2019Terminated
Potential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.
Z-0128-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 29, 2019Terminated
Potential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.
Z-0127-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 29, 2019Terminated
Potential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.
Z-0126-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported Plum problems
Device problems
- Fluid/Blood Leak875
- No Flow148
- Crack68
- Break56
- Particulates53
- Air/Gas in Device41
- Device Contamination with Chemical or Other Material38
- Contamination33
Patient problems
- No Clinical Signs, Symptoms or Conditions1,122
- No Consequences Or Impact To Patient191
- Chemical Exposure62
- No Patient Involvement14
- Insufficient Information7
- No Known Impact Or Consequence To Patient6
- Discomfort3
- Low Blood Pressure/ Hypotension3
Plum models
26 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Plum 1468728Set, administration, intravascular1 recall2 deaths482 reports
- Plum 1425628Set, administration, intravascular1 recall204 reports
- Plum 1425528Set, administration, intravascular1 recall164 reports
- Plum 1427329Set, administration, intravascular214 reports
- Plum 1233912Set, administration, intravascular1 recall131 reports
- Plum 1187912Pump, infusion1 recall8 reports
- Plum 1425128Set, administration, intravascular1 recall1 report
- Plum 1495388Set, administration, intravascular1 recall0 reports
- Plum 1973201Set, administration, intravascular1 recall0 reports
- Plum 1427128Set, administration, intravascular1 recall0 reports
- Plum 1467928Set, administration, intravascular1 recall0 reports
- Plum 1495588Set, administration, intravascular1 recall0 reports
- Plum 2061928Set, administration, intravascular1 recall0 reports
- Plum 1424328Set, administration, intravascular1 recall0 reports
- Plum 2400701Set, administration, intravascular1 recall0 reports
- Plum 1424828Set, administration, intravascular43 reports
- Plum 1253905Set, administration, intravascular41 reports
- Plum 1425428Set, administration, intravascular36 reports
- Plum 1421228Set, administration, intravascular13 reports
- Plum 1424228Set, administration, intravascular6 reports
- Plum 2079225Pump, infusion4 reports
- Plum 1194802Set, administration, intravascular3 reports
- Plum 1203012Set, administration, intravascular3 reports
- Plum 1239125Pump, infusion3 reports
- Plum 1253828Set, administration, intravascular1 report
- Plum 1955488Set, administration, intravascular1 report
Recent Plum adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951774 ↗MalfunctionSep 30, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951595 ↗MalfunctionSep 29, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937656 ↗MalfunctionSep 29, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936116 ↗MalfunctionSep 29, 2026Fluid/Blood Leak; Free or Unrestricted Flow; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934694 ↗MalfunctionSep 28, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50924633 ↗MalfunctionSep 28, 2026Environmental Particulates
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50924374 ↗MalfunctionSep 28, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923546 ↗MalfunctionSep 28, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923251 ↗MalfunctionSep 25, 2026Obstruction of Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50903120 ↗MalfunctionSep 25, 2026Environmental Particulates
The event involved a Primary Plum Set reported that there is dirt in the drip chamber. The issue was noted during priming. There were no concomitant drug and concomitant device involved. There was no bleedback, no biohazard, no user facility MedWatch, and no chemo. The device was not reprocessed or resterilized.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50897908 ↗MalfunctionSep 24, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50888192 ↗MalfunctionSep 24, 2026Free or Unrestricted Flow
The event involved a Primary PLUMSET, PE Lined Tubing, 107 Inch. It was reported that even after pressing the stop button, the medicine keeps flowing and won't stop. This issue led to over delivery during infusion. There was patient involvement but no patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50883060 ↗MalfunctionSep 23, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873863 ↗MalfunctionSep 23, 2026Break; Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873245 ↗
Other ICU Medical, Inc. products
- CADD123 models · 53,012 reports · 21 recalls
- Medfusion16 models · 13,114 reports · 5 recalls
- Bivona469 models · 2,811 reports · 29 recalls
- Level 152 models · 3,733 reports · 1 recall
- Portex409 models · 3,479 reports · 5 recalls
- Cleo6 models · 2,782 reports
- Pneupac23 models · 1,972 reports
- Medex98 models · 774 reports · 23 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page