Medline ReNewal HAREP103635Recalls & reported problems
Medline Industries, Inc. · 375-544-000 @FRML AGGRESSIVE PLS CTR LG · Class 2
All models of Medline ReNewal (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10888277379855), received between Nov 1, 2016 and Jun 2, 2017 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Fragmentation4
- Electrical /Electronic Property Problem1
Patient problems
- No Known Impact Or Consequence To Patient5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJun 2, 2017Material Fragmentation; Material Fragmentation
WE RECEIVED A REPORT INDICATING THAT THE TIP OF A REPROCESSED ARTHROSCOPIC SHAVER DETACHED DURING USE. THE PHYSICIAN WAS ABLE TO LOCATE AND REMOVE THE DETACHED TIP WITH A GRASPER. A NEW DEVICE AND A BAG OF FLUID WERE NEEDED TO COMPLETE THE CASE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6607354 ↗InjuryFeb 3, 2017Material Fragmentation; Material Fragmentation
WE RECEIVED A REPORT INDICATING THAT A PIECE OF A REPROCESSED ARTHROSCOPIC SHAVER DISLODGED AND FELL INTO THE PATIENT DURING USE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6301703 ↗MalfunctionNov 8, 2016Material Fragmentation; Material Fragmentation
MEDLINE RENEWAL RECEIVED A REPORT INDICATING THAT THE TIP OF THE REPROCESSED STRYKER FORMULA AGGRESSIVE PLUS CUTTER, MODEL 375-544-000, BROKE OFF IN THE PATIENT'S SHOULDER DURING USE. THE TIP WAS RETRIEVED BY THE SURGEON, AND DID NOT RESULT IN PATIENT HARM. ANOTHER DEVICE WAS READILY AVAILABLE TO COMPLETE THE PROCEDURE; THEREFORE NO ADDITIONAL MEDICAL INTERVENTION WAS REPORTED AS A RESULT OF THE INCIDENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6088246 ↗MalfunctionNov 8, 2016Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
MEDLINE RENEWAL RECEIVED A REPORT INDICATING A REPROCESSED DEPUY MITEK VAPR SIDE EFFECT ELECTRODE, W/INTEGRATED HAND PIECE, MODEL 227301, WAS NOT RECOGNIZED BY THE GENERATOR WHEN IT WAS PLUGGED IN DURING A SURGICAL PROCEDURE. THEREFORE THE DEVICE COULD NOT BE USED. THE REPORT DID NOT INDICATE THAT THE DEVICE FAILURE RESULTED IN ANY PATIENT HARM. ANOTHER DEVICE WAS READILY AVAILABLE TO COMPLETE THE PROCEDURE; THEREFORE NO PATIENT HARM OR ADDITIONAL MEDICAL INTERVENTION WAS REPORTED AS A RESULT OF THE INCIDENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6088216 ↗MalfunctionNov 1, 2016Material Fragmentation; Material Fragmentation
MEDLINE RENEWAL RECEIVED A REPORT INDICATING THAT THE TIP OF A REPROCESSED STRYKER FORMULA AGGRESSIVE PLUS CUTTER, MODEL 375-544-000, BROKE OFF IN THE PATIENT'S SHOULDER DURING USE. THE REPORT DID NOT INDICATE THAT THE DEVICE FAILURE RESULTED IN PATIENT HARM. ANOTHER DEVICE WAS READILY AVAILABLE TO COMPLETE THE PROCEDURE; THEREFORE NO PATIENT HARM OR ADDITIONAL MEDICAL INTERVENTION WAS REPORTED AS A RESULT OF THE INCIDENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6071501 ↗
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page