Skip to content

CHESAPEAKE® Stabilization System 2013-21530L5Recalls & reported problems

VB Spine LLC · Implant · Class 2 · Implant

All models of CHESAPEAKE Stabilization System (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10888857027886), received between Mar 2, 2022 and Mar 2, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Misassembled” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Misassembled1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 2, 2022Misassembled; Misassembled

    IT WAS REPORTED THAT DURING A CHESAPEAKE LUMBER CASE, THE UNIVERSAL AWL 'SKIVED' OFF THE ENDPLATE AND DID NOT CREATE THE NECESSARY PILOT HOLE FOR THE SCREW. THIS RESULTED IN THE CAGES AT S1 AND L4 HAVING ONLY 2 SCREWS INSERTED INSTEAD OF 3. ADDITIONALLY, DURING SCREW INSERTION, THE ANGLED TAPERED DRIVER CAME OFF OF THE SCREW. IT WAS ALSO OBSERVED THAT THE LASER MARKINGS HAD WORN OFF OF THE EXPANDABLE DRILL GUIDE. THIS REPORT WILL CAPTURE THE SECOND OF TWO CAGES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13655640 ↗

Other Intervertebral fusion device with integrated fixation, lumbar devices

All Intervertebral fusion device with integrated fixation, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page