Skip to content

EVEREST® Spinal System 2911-06545Recalls & reported problems

VB Spine LLC · Polyaxial Screw Size Ø6.5x45 mm · Class 2 · Implant

All models of EVEREST Spinal System (77)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10888857042339), received between Apr 5, 2017 and Jun 25, 2020 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Material Deformation1

Patient problems

  • Injury1
  • No Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 25, 2020Material Deformation; Material Deformation

    IT WAS REPORTED THAT WHILE TRYING TO PLACE A TEMPORARY ROD IN PREPARATION FOR TLIF, AN EVEREST MI SINGLE ACTION ROD REDUCER TIP FRACTURED AND BECAME STUCK ON AN EVEREST POLAYAXIAL SCREW AT THE L5 LEVEL. THE SURGEON WAS ABLE TO GET THE REDUCER OFF OF THE SCREW WITH AGGRESSIVE FORCE. AFTER TLIF, THE SURGEON TRIED TO REDUCE THE SAME LEFT L5 SCREW WHILE ATTEMPTING TO PLACE THE FINAL RODS. BOTH BASECAMP ROD REDUCERS POPPED OFF OF THE SCREW ONCE THEY WERE FULLY REDUCED. THE SURGEON THEN TRIED TO USE AN EVEREST MI OFFSET ROD REDUCER, BUT IT ALSO POPPED OFF. IT WAS DETERMINED THAT THE SCREW WAS BENT AND IT WAS REMOVED AND REPLACED WITH A NEW ONE. SURGERY WAS SUCCESSFULLY COMPLETED WITH ANOTHER REDUCER AND AN HOUR AND A HALF DELAY. THIS REPORT REPRESENTS THE SCREW.

    Patient problems: Injury

    MAUDE report 10197603 ↗
  • InjuryApr 5, 2017Break; Break

    ON (B)(6) 2017, IT WAS REPORTED TO K2M, INC. THAT A PATIENT PRESENTED WITH A POSSIBLE SCREW BREAK. THE PATIENT REPORTEDLY HAD POSSIBLE SCREW BREAK APPROXIMATELY 5 MONTHS POST-OP. THERE ARE NO PLANS TO REVISE AT THIS TIME.

    Patient problems: No Information

    MAUDE report 6460280 ↗

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page