EVEREST Spinal System 5101-90218Recalls & reported problems
VB Spine LLC · XT Support Tower · Class 1
All models of EVEREST Spinal System (77)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10888857325968), received between Nov 30, 2020 and Mar 21, 2023 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Mechanical Jam6
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Bone Fracture(s)1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2023Mechanical Jam; Mechanical Jam
A COMPANY REPRESENTATIVE REPORTED THAT DURING A SCOLIOSIS CORRECTION CASE AN EVEREST XT SUPPORT TOWER AND AN EVEREST XT SUPPORT TOWER REDUCER JAMMED / BECAME STUCK TOGETHER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 10 MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT IS FOR THE EVEREST XT SUPPORT TOWER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16587171 ↗MalfunctionMar 2, 2023Mechanical Jam; Mechanical Jam
A COMPANY REPRESENTATIVE REPORTED THAT DURING A SCOLIOSIS CORRECTION CASE, AN EVEREST XT SUPPORT TOWER JAMMED AND BECAME STUCK IN AN EVEREST XT SUPPORT TOWER REDUCER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 10 MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT IS FOR THE EVEREST XT SUPPORT TOWER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16469098 ↗MalfunctionFeb 24, 2023Mechanical Jam; Mechanical Jam
A COMPANY REPRESENTATIVE REPORTED THAT DURING A SCOLIOSIS CORRECTION CASE TWO EVEREST XT SUPPORT TOWERS JAMMED / GOT STUCK IN TWO EVEREST XT SUPPORT TOWER REDUCERS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 15 MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT IS FOR THE SECOND OF TWO EVEREST XT SUPPORT TOWERS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16437628 ↗MalfunctionFeb 24, 2023Mechanical Jam; Mechanical Jam
A COMPANY REPRESENTATIVE REPORTED THAT DURING A SCOLIOSIS CORRECTION CASE TWO EVEREST XT SUPPORT TOWERS JAMMED / GOT STUCK IN TWO EVEREST XT SUPPORT TOWER REDUCERS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 15 MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT IS FOR THE FIRST OF TWO EVEREST XT SUPPORT TOWERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16437615 ↗MalfunctionAug 17, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT AN EVEREST XT SUPPORT TOWER WAS JAMMING INTRA-OPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12333519 ↗InjuryNov 30, 2020Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING INTRA-OPERATIVE ROD REDUCTION A SCREW BECAME STUCK IN AN EVEREST XT SUPPORT TOWER AND EVEREST XT SUPPORT TOWER REDUCER ASSEMBLY CAUSING THE PEDICLE AT L5 TO BREAK. THE SURGEON WAS THEN UNABLE TO PLACE A SCREW AT THE RIGHT L5 LEADING TO ONLY UNILATERAL POSTERIOR FIXATION BEING ACHIEVED. THIS SUBSEQUENTLY LED TO AN APPROXIMATE 2-3 HOUR SURGICAL DELAY. THIS RECORD REPRESENTS THE XT SUPPORT TOWER.
Patient problems: Bone Fracture(s)
MAUDE report 10924219 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page