Skip to content

YUKON OCT Spinal System 7601-03514Recalls & reported problems

VB Spine LLC · Polyaxial Screw Ø3.5x14 mm · Class 2 · Implant

All models of YUKON OCT Spinal System (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10888857349315), received between Aug 2, 2019 and Jan 2, 2024 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • Migration2
  • Physical Resistance/Sticking2
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1

Patient problems

  • Injury2
  • No Clinical Signs, Symptoms or Conditions2
  • Neuropathy1
  • No Information1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 2, 2024Insufficient Device Problem Information; Migration; Insufficient Device Problem Information; Migration

    A PATIENT EXPERIENCED MIGRATION OF A C7 YUKON SET SCREW APPROXIMATELY TWO MONTHS AFTER IMPLANTATION. REVISION SURGERY IS NOT CURRENTLY PLANNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18428103 ↗
  • MalfunctionDec 20, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT RECEIVED THEIR INITIAL SURGERY ON (B)(6) 2020 WHERE THEY RECEIVED A POSTERIOR CERVICAL FIXATION / FUSION. THE PATIENT THEN PRESENTED WITH CONSISTENT MYELOPATHY AMONG OTHER PAIN COMPLAINTS IN THEIR CERVICAL AREA. AN EXTENSION OF THE POSTERIOR CERVICAL LATERAL MASS FUSION AS WELL LAMINECTOMY WAS THEN PERFORMED IN (B)(6) 2021. DURING THE PLANNED REVISION SURGERY, THE YUKON SCREW AT C5 WHICH WAS ORIGINALLY IMPLANTED IN (B)(6) 2020 WAS NOTICED TO BE FRACTURED DURING REMOVAL. HALF OF THE SHAFT ATTACHED TO THE SCREW HEAD WAS ABLE TO BE REMOVED WHILE THE OTHER HALF REMAINS IN THE PATIENT. WELL ESTABLISHED BONE FUSION WAS NOTED ON THE REMAINING IMPLANTED SCREWS.

    Patient problems: Neuropathy; No Clinical Signs, Symptoms or Conditions; Pain

    MAUDE report 13047668 ↗
  • InjuryOct 8, 2019Adverse Event Without Identified Device or Use Problem; Physical Resistance/Sticking; Adverse Event Without Identified Device or Use Problem

    A YUKON SCREW INSERTER "STUCK" TO THE SCREW WHILE INSERTING AT THE RIGHT C3, CAUSING THE SURGEON TO APPLY ADDITIONAL FORCE. THIS FORCE BROKE THROUGH THE LATERAL MASS WHICH MADE IT ESSENTIAL TO EXTEND THE CONSTRUCT UP TO THE RIGHT C2 RESULTING IN A 10 MINUTE SURGICAL DELAY. THIS RECORD CAPTURES THE POLYAXIAL SCREW . IT WAS DETERMINED THAT THE YUKON OCT POLYAXIAL SCREW HAD NO CAUSE TO THE ADVERSE EVENT AND THAT THE DEVICE DID NOT MALFUNCTION. THEREFORE, THIS DEVICE IS A CONCOMITANT PRODUCT.

    Patient problems: Injury

    MAUDE report 9169098 ↗
  • MalfunctionAug 30, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPANY REPRESENTATIVE REPORTED THAT THE SCREW SITE PREPARATION, THE YUKON POLYAXIAL SCREW Ø3.5X14 MM WAS "DIFFICULT TO LOAD SECURELY THE INSERT DRIVER" THE EVENT WAS NOTED DURING A CASE PREPARATION; THERE WAS NO PATIENT INVOLVEMENT. THE CASE WAS SUCCESSFULLY EXECUTED USING ALTERNATIVE SCREWS THAT WERE READILY AVAILABLE TO THE PHYSICIAN. THERE WAS NO HARM TO THE PATIENT AS THE EVENT WAS IDENTIFIED PREOPERATIVELY.

    Patient problems: No Information

    MAUDE report 8950626 ↗
  • InjuryAug 2, 2019Migration; Migration; Physical Resistance/Sticking

    A PHYSICIAN REPORTED THAT A YUKON POLYAXIAL SCREW BECAME ¿STUCK¿ AT THE HEAD POSTOPERATIVELY. FOLLOWING IMPLANT SURGERY, THE PATIENT DID NOT WEAR THEIR CERVICAL COLLAR AS DIRECTED BY THE PHYSICIAN AND, ¿WENT BACK TO REGULAR ACTIVITIES BEFORE THEY WERE ADVISED TO.¿ THE PATIENT WAS REVISED AND LAMINA SCREWS C3 THROUGH C6 WERE REPLACED. PATIENT WAS REVISED AND ALL LAMINA SCREWS C3-C6 WERE REPLACED.

    Patient problems: Injury

    MAUDE report 8857535 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates