YUKON OCT Spinal System 7601-04526Recalls & reported problems
VB Spine LLC · Polyaxial Screw Ø4.5x26 mm · Class 2 · Implant
All models of YUKON OCT Spinal System (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10888857349797), received between Dec 20, 2021 and Feb 26, 2026 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Fracture” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture8
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 26, 2026Fracture
A COMPANY REPRESENTATIVE REPORTED THAT THE NECK OF A YUKON POLYAXIAL SCREW FRACTURED POST-OPERATIVELY. REVISION SURGERY HAS NOT BEEN PERFORMED NOR SCHEDULED AT THIS TIME.
Patient problems: Insufficient Information
MAUDE report 24455577 ↗MalfunctionDec 22, 2024Fracture; Fracture
IT WAS REPORTED THAT A YUKON OCT POLYAXIAL SCREW FRACTURED INTRA-OPERATIVELY, THE SCREW WAS UNABLE TO BE REMOVED AND LEFT IN PEDICLE WITHOUT TULIP HEAD. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 45-MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20997697 ↗MalfunctionDec 17, 2024Fracture; Fracture
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20958934 ↗MalfunctionDec 6, 2024Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT ONE YUKON POLYAXIAL SCREW FRACTURED AT T2 AND ONE YUKON SET SCREW MIGRATED APPROXIMATELY 6 MONTHS POST-OPERATIVELY. REVISION SURGERY HAS NOT BEEN PERFORMED. THIS REPORT CAPTURES THE POLYAXIAL SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20865231 ↗MalfunctionDec 6, 2024Fracture; Migration; Fracture; Migration
A COMPANY REPRESENTATIVE REPORTED THAT ONE YUKON POLYAXIAL SCREW FRACTURED AT T2 AND ONE YUKON SET SCREW MIGRATED APPROXIMATELY 6 MONTHS POST-OPERATIVELY. REVISION SURGERY HAS NOT BEEN PERFORMED. THIS REPORT CAPTURES THE SET SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20865204 ↗InjuryJun 27, 2024Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY 2 MONTHS POST-OPERATIVELY TO ADDRESS A FRACTURED YUKON POLYAXIAL SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19623791 ↗MalfunctionJul 29, 2022Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF Q2 2022 RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. THIS REPORT CAPTURES ONE INSTANCE OF INTRAOPERATIVE YUKON POLYAXIAL SCREW FRACTURE. THE ASSOCIATED PROCEDURE WAS COMPLETED SUCCESSFULLY WITH AN APPROXIMATELY THIRTY-MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THE SHAFT OF THE FRACTURED SCREW REMAINS IN THE PATIENT'S BONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15135614 ↗MalfunctionDec 20, 2021Fracture; Fracture
IT WAS REPORTED THAT THE HEAD OF A YUKON POLYAXIAL 'BROKE FROM BETWEEN HEAD AND PROXIMAL SHAFT' INTRA-OPERATIVELY. SURGERY WAS SUCCESSFULLY COMPLETED WITH AN HOUR DELAY. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13043197 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page