YUKON OCT Spinal System 7601-05024Recalls & reported problems
VB Spine LLC · Polyaxial Screw Ø5.0x24 mm · Class 2 · Implant
All models of YUKON OCT Spinal System (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10888857349995), received between Jun 26, 2020 and Dec 4, 2024 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture7
- Material Separation1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryDec 4, 2024Fracture; Fracture
IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS TWO YUKON POLYAXIAL SCREWS THAT FRACTURED APPROXIMATELY ONE MONTH POST-OPERATIVELY. THIS REPORT CAPTURES THE FIRST OF TWO YUKON POLYAXIAL SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20846664 ↗InjuryDec 4, 2024Fracture; Fracture
IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS TWO YUKON POLYAXIAL SCREWS THAT FRACTURED APPROXIMATELY ONE MONTH POST-OPERATIVELY. THIS REPORT CAPTURES THE SECOND OF TWO YUKON POLYAXIAL SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20846521 ↗InjuryMar 23, 2021Fracture; Fracture
IT WAS REPORTED THAT TWO YUKON OCT POLYAXIAL SCREWS BROKE THREE WEEKS POST-OPERATIVELY. THE PATIENT IS REPORTED TO HAVE "FELT THE SCREWS DISLODGE" WHILE OPENING A HEAVY DOOR. REVISION HAS OCCURRED. THIS REPORT WILL REPRESENT THE FIRST OF TWO SCREWS.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11555925 ↗MalfunctionMar 15, 2021Fracture; Fracture
IT WAS REPORTED THAT 2 YUKON OCT POLYAXIAL SCREWS AT T1 BROKE POST-OPERATIVELY. REVISION SURGERY HAS NOT BEEN SCHEDULED AT THIS TIME. THIS REPORT WILL CAPTURE THE SECOND OF TWO SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11490064 ↗MalfunctionMar 15, 2021Fracture; Fracture
IT WAS REPORTED THAT 2 YUKON OCT POLYAXIAL SCREWS AT T1 BROKE POST-OPERATIVELY. REVISION SURGERY HAS NOT BEEN SCHEDULED AT THIS TIME. THIS REPORT WILL CAPTURE THE FIRST OF TWO SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11489952 ↗MalfunctionDec 8, 2020Nonstandard Device; Nonstandard Device
A STRYKER SALES REPRESENTATIVE IDENTIFIED AN INCORRECTLY MARKED YUKON POLYAXIAL SCREW DURING RESTOCK OF A CONSIGNMENT SET TRAY, OUTSIDE THE OPERATING THEATER. THE SUBJECT SCREW WAS INCORRECTLY LASER MARKED WITH SIZE Ø5.0 (DIAMETER) & 24 (MM LENGTH) WITH CATALOG NUMBER 7601-05024. HOWEVER, THE SCREW SHOULD HAVE BEEN MARKED Ø4.0 (DIAMETER) & 26 (MM LENGTH) WITH ASSOCIATED CATALOG NUMBER 7601-04026. THERE WAS NO ASSOCIATED PROCEDURE AND THEREFORE NO ADVERSE CONSEQUENCE TO A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10969239 ↗InjurySep 23, 2020Fracture; Material Separation; Fracture; Material Separation
IT WAS REPORTED THAT DURING INTRA-OPERATIVE SCREW INSERTION INTO THE T1 PEDICLE A YUKON POLYAXIAL SCREW FRACTURED AT THE SHAFT. THE SCREW SHAFT COULD NOT BE REMOVED AND REMAINS AT THAT LEVEL. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 5-10 MINUTE SURGICAL DELAY BY BYPASSING THAT LEVEL AND PROCEEDING ON WITH THE CONSTRUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10574761 ↗MalfunctionJun 26, 2020Fracture; Fracture
IT WAS REPORTED THAT THE HEAD OF A YUKON OCT SPINAL SYSTEM POLYAXIAL SCREW BROKE OFF OF THE SHANK DURING INSERTION INTO THE PEDICLE OF T2. THE SCREW WAS REMOVED AND SURGERY WAS SUCCESSFULLY COMPLETED WITH NO DELAY. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10203370 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page