Skip to content

YUKON OCT Spinal System 7601-04020MRecalls & reported problems

VB Spine LLC · Polyaxial Screw ML Ø4.0x20 mm · Class 2 · Implant

All models of YUKON OCT Spinal System (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10888857350397), received between Jun 9, 2021 and Jun 9, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2021Fracture; Fracture

    IT WAS REPORTED THAT TWO YUKON POLYAXIAL SCREWS FRACTURED AT T1, ABOUT 5 MONTHS POST-OPERATIVELY. NO ADVERSE CONSEQUENCES WERE REPORTED. REVISION SURGERY HAS NOT BEEN PERFORMED NOR SCHEDULED AT THIS TIME. THIS RECORD CAPTURES THE SECOND OF THE TWO REPORTED YUKON POLYAXIAL SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11967763 ↗
  • MalfunctionJun 9, 2021Fracture; Fracture

    IT WAS REPORTED THAT TWO YUKON POLYAXIAL SCREWS FRACTURED AT T1, ABOUT 5 MONTHS POST- OPERATIVELY. NO ADVERSE CONSEQUENCES WERE REPORTED. REVISION SURGERY HAS NOT BEEN PERFORMED NOR SCHEDULED AT THIS TIME. THIS RECORD CAPTURES THE FIRST OF THE TWO REPORTED YUKON POLYAXIAL SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11967742 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page