Skip to content

Ventlab AF2140MBPRecalls & reported problems

Ventlab, LLC · AirFlow Pediatric Resuscitator with Integrated Manometer, B/V Filter, Bag Reservoir, Mask, and PEEP Valve · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889483088111), received between May 26, 2023 and Jan 26, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 11, 2023Open, Classified

    A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

    Z-0361-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Material Integrity Problem1
  • Pressure Problem1

Patient problems

  • Decreased Level of Consciousness1
  • Unspecified Respiratory Problem1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 26, 2024Pressure Problem; Pressure Problem

    PATIENT PRESENTATION: AN UNRESPONSIVE PATIENT, HELMETED DRIVER OF A MOTORCYCLE STRUCK BY ANOTHER VEHICLE AT A HIGH RATE OF SPEED THEN STRUCK AND RUN OVER BY AN ADDITIONAL VEHICLE, WAS ASSESSED WITH A GLASGOW COMA SCALE (GCS) SCORE OF 3, INDICATIVE OF SEVERE CONSCIOUSNESS IMPAIRMENT. MEDICAL INTERVENTIONS: RAPID SEQUENCE INTUBATION (RSI) WAS PERFORMED TO ESTABLISH AIRWAY PATENCY. THE ENDOTRACHEAL TUBE (ETT) WAS VENTILATED WITH AN ADULT SUNMED (AF2140MBP) MANUAL RESUSCITATOR WITH A MANOMETER, POSITIVE END-EXPIRATORY PRESSURE (PEEP) VALVE SET AT 5 CMH2O, HEAT AND MOISTURE EXCHANGE FILTER (HMFE), IN-LINE END-TIDAL CO2 (ETCO2) SENSOR AND CONNECTED TO A 15 L/MIN OXYGEN SOURCE. FIC02 ELEVATION ISSUE: WHILE USING THE SUNMED MANUAL RESUSCITATOR, THE PATIENT'S FRACTION OF INSPIRED CO2 (FICO2) LEVELS INCREASED UNTIL THE CLINICAL TEAM PLACED THE PATIENT ON A VENTILATOR. THIS INCREASE BEGAN AT 14:38.36 (FIGURE 1) AND PEAKED AT 20 MMHG AT 14:42.13 (FIGURE 2). FICO2 LEVELS FLUCTUATED BETWEEN 10 AND 20 MMHG (FIGURE 3) UNTIL THE PATIENT WAS TRANSITIONED TO A VENTILATOR AT 14:49.51. THE ELEVATION IN FICO2 RESOLVED ONCE THE VENTILATOR WAS INITIATED (FIGURE 4). PRODUCT PROBLEM: SUSPICIONS ARISE THAT T

    Patient problems: Decreased Level of Consciousness

    MAUDE report 18592409 ↗
  • MalfunctionMay 26, 2023Material Integrity Problem; Material Integrity Problem

    BAGS AREN'T HOLDING PRESSURE.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 17014849 ↗

Other Ventilator, Emergency, Manual (Resuscitator) devices

All Ventilator, Emergency, Manual (Resuscitator) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.