Ventlab AF2140MBPRecalls & reported problems
Ventlab, LLC · AirFlow Pediatric Resuscitator with Integrated Manometer, B/V Filter, Bag Reservoir, Mask, and PEEP Valve · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889483088111), received between May 26, 2023 and Jan 26, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Integrity Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Oct 11, 2023Class IIOpen, Classified
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Z-0361-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Integrity Problem1
- Pressure Problem1
Patient problems
- Decreased Level of Consciousness1
- Unspecified Respiratory Problem1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 26, 2024Pressure Problem; Pressure Problem
PATIENT PRESENTATION: AN UNRESPONSIVE PATIENT, HELMETED DRIVER OF A MOTORCYCLE STRUCK BY ANOTHER VEHICLE AT A HIGH RATE OF SPEED THEN STRUCK AND RUN OVER BY AN ADDITIONAL VEHICLE, WAS ASSESSED WITH A GLASGOW COMA SCALE (GCS) SCORE OF 3, INDICATIVE OF SEVERE CONSCIOUSNESS IMPAIRMENT. MEDICAL INTERVENTIONS: RAPID SEQUENCE INTUBATION (RSI) WAS PERFORMED TO ESTABLISH AIRWAY PATENCY. THE ENDOTRACHEAL TUBE (ETT) WAS VENTILATED WITH AN ADULT SUNMED (AF2140MBP) MANUAL RESUSCITATOR WITH A MANOMETER, POSITIVE END-EXPIRATORY PRESSURE (PEEP) VALVE SET AT 5 CMH2O, HEAT AND MOISTURE EXCHANGE FILTER (HMFE), IN-LINE END-TIDAL CO2 (ETCO2) SENSOR AND CONNECTED TO A 15 L/MIN OXYGEN SOURCE. FIC02 ELEVATION ISSUE: WHILE USING THE SUNMED MANUAL RESUSCITATOR, THE PATIENT'S FRACTION OF INSPIRED CO2 (FICO2) LEVELS INCREASED UNTIL THE CLINICAL TEAM PLACED THE PATIENT ON A VENTILATOR. THIS INCREASE BEGAN AT 14:38.36 (FIGURE 1) AND PEAKED AT 20 MMHG AT 14:42.13 (FIGURE 2). FICO2 LEVELS FLUCTUATED BETWEEN 10 AND 20 MMHG (FIGURE 3) UNTIL THE PATIENT WAS TRANSITIONED TO A VENTILATOR AT 14:49.51. THE ELEVATION IN FICO2 RESOLVED ONCE THE VENTILATOR WAS INITIATED (FIGURE 4). PRODUCT PROBLEM: SUSPICIONS ARISE THAT T
Patient problems: Decreased Level of Consciousness
MAUDE report 18592409 ↗MalfunctionMay 26, 2023Material Integrity Problem; Material Integrity Problem
BAGS AREN'T HOLDING PRESSURE.
Patient problems: Unspecified Respiratory Problem
MAUDE report 17014849 ↗
FDA clearance and classification
The FDA regulates product code BTM (ventilator, emergency, manual (resuscitator)) under 21 CFR 868.5915 as a Class II device. 21 CFR 868.5915 ↗
Devices of this type normally need 510(k) clearance before sale.
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Ventlab AF2140MBP
Has the Ventlab AF2140MBP been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0361-2024 (Class II), began on Oct 11, 2023: A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Ventlab AF2140MBP?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between May 26, 2023 and Jan 26, 2024. In those reports, the most common specific device problems are “Material Integrity Problem” (1) and “Pressure Problem” (1); the most common patient problems are “Decreased Level of Consciousness” (1) and “Unspecified Respiratory Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Ventlab AF2140MBP?
None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Ventlab AF2140MBP, and is it still sold?
The FDA's device database lists Ventlab, LLC as the company responsible for it (brand Ventlab, model AF2140MBP). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Ventilator, Emergency, Manual (Resuscitator) devices
- Ventlab AF5140MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall8 reports
- Curaplex AF2000Bound Tree Medical, LLC · Ventilator, emergency, manual (resuscitator)1 recall7 reports
- Ventlab AF5140MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall5 reports
- Ventlab AF5142MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall2 reports
- Ventlab SS3200MB-MMCVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Medline CPRM1116FMMedline Industries, Inc. · Ventilator, emergency, manual (resuscitator)1 recall1 report
- Medline CPRM1126FPWMMedline Industries, Inc. · Ventilator, emergency, manual (resuscitator)1 recall1 report
- Ventlab AF1140MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.