Ventlab AF2140MBPRecalls & reported problems
Ventlab, LLC · AirFlow Pediatric Resuscitator with Integrated Manometer, B/V Filter, Bag Reservoir, Mask, and PEEP Valve · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889483088111), received between May 26, 2023 and Jan 26, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Integrity Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 11, 2023Open, Classified
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Z-0361-2024 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Material Integrity Problem1
- Pressure Problem1
Patient problems
- Decreased Level of Consciousness1
- Unspecified Respiratory Problem1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 26, 2024Pressure Problem; Pressure Problem
PATIENT PRESENTATION: AN UNRESPONSIVE PATIENT, HELMETED DRIVER OF A MOTORCYCLE STRUCK BY ANOTHER VEHICLE AT A HIGH RATE OF SPEED THEN STRUCK AND RUN OVER BY AN ADDITIONAL VEHICLE, WAS ASSESSED WITH A GLASGOW COMA SCALE (GCS) SCORE OF 3, INDICATIVE OF SEVERE CONSCIOUSNESS IMPAIRMENT. MEDICAL INTERVENTIONS: RAPID SEQUENCE INTUBATION (RSI) WAS PERFORMED TO ESTABLISH AIRWAY PATENCY. THE ENDOTRACHEAL TUBE (ETT) WAS VENTILATED WITH AN ADULT SUNMED (AF2140MBP) MANUAL RESUSCITATOR WITH A MANOMETER, POSITIVE END-EXPIRATORY PRESSURE (PEEP) VALVE SET AT 5 CMH2O, HEAT AND MOISTURE EXCHANGE FILTER (HMFE), IN-LINE END-TIDAL CO2 (ETCO2) SENSOR AND CONNECTED TO A 15 L/MIN OXYGEN SOURCE. FIC02 ELEVATION ISSUE: WHILE USING THE SUNMED MANUAL RESUSCITATOR, THE PATIENT'S FRACTION OF INSPIRED CO2 (FICO2) LEVELS INCREASED UNTIL THE CLINICAL TEAM PLACED THE PATIENT ON A VENTILATOR. THIS INCREASE BEGAN AT 14:38.36 (FIGURE 1) AND PEAKED AT 20 MMHG AT 14:42.13 (FIGURE 2). FICO2 LEVELS FLUCTUATED BETWEEN 10 AND 20 MMHG (FIGURE 3) UNTIL THE PATIENT WAS TRANSITIONED TO A VENTILATOR AT 14:49.51. THE ELEVATION IN FICO2 RESOLVED ONCE THE VENTILATOR WAS INITIATED (FIGURE 4). PRODUCT PROBLEM: SUSPICIONS ARISE THAT T
Patient problems: Decreased Level of Consciousness
MAUDE report 18592409 ↗MalfunctionMay 26, 2023Material Integrity Problem; Material Integrity Problem
BAGS AREN'T HOLDING PRESSURE.
Patient problems: Unspecified Respiratory Problem
MAUDE report 17014849 ↗
Other Ventilator, Emergency, Manual (Resuscitator) devices
- Ventlab AF5140MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall8 reports
- Curaplex AF2000Bound Tree Medical, LLC · Ventilator, emergency, manual (resuscitator)1 recall6 reports
- Ventlab AF5140MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall5 reports
- Ventlab AF5142MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall2 reports
- Ventlab AF1140MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Ventlab AF2142MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 death1 report
- Ventlab VN3100OBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Medline CPRM1126FPWMMedline Industries, Inc. · Ventilator, emergency, manual (resuscitator)1 recall1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.