Ventlab AF5140MBRecalls & reported problems
Ventlab, LLC · AirFlow Adult Resuscitator with Integrated Manometer, B/V Filter, Bag, Reservoir, and Mask · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10889483103012), received between Oct 18, 2018 and Aug 15, 2025 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Defective Component” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 11, 2023Open, Classified
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Z-0361-2024 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Defective Component2
- Adverse Event Without Identified Device or Use Problem1
- Component Missing1
- Difficult to Open or Close1
- Disconnection1
- Gas/Air Leak1
- Incomplete or Inadequate Connection1
- Loose or Intermittent Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Decreased Respiratory Rate2
- Insufficient Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 15, 2025Component Missing; Component Missing
IT WAS REPORTED, DURING A CODE BLUE EVENT IN THE CARDIAC PROGRESSIVE CARE UNIT 2E, THE MEDICAL TEAM OPENED THE KIT TO OBTAIN THE AIRFLOW BAG AND DISCOVERED ONLY THE RESUSCITATION BAG WAS IN THE KIT; THE MASK WAS MISSING. ANOTHER AIRFLOW BAG WAS OBTAINED; THE EVENT CAUSED A DELAY IN TREATMENT; HOWEVER, NO PATIENT HARM OR INJURY OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22819447 ↗MalfunctionMar 4, 2025Misassembled; Misassembled
IT WAS REPORTED, DIRECTLY AFTER INTUBATION THE ANESTHESIOLOGIST COULD NOT CONNECT THE AMBU-BAG TO THE ENDOTRACHEAL TUBE (ETT). IT APPEARED THE DEVICE WAS INCORRECTLY ASSEMBLED; THERE WAS NO REPORTED INJURY TO THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 21520852 ↗MalfunctionFeb 16, 2024Disconnection; Disconnection
THE END OF THE RESUSCITATION BAG POPPED OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18727884 ↗MalfunctionSep 21, 2023Gas/Air Leak; Gas/Air Leak
AIR LEAKING DURING INTUBATION, NOT DELIVERING THE PATIENT O2. OCCURRED DURING PATIENT USE, POSSIBLE DELAY IN TREATMENT.
Patient problems: Decreased Respiratory Rate
MAUDE report 17795871 ↗InjuryJan 6, 2023Loose or Intermittent Connection; Incomplete or Inadequate Connection; Loose or Intermittent Connection; Incomplete or Inadequate Connection
PATIENT DISCONNECTED FROM THE VENTILATOR AND PLACED ON AMBU BAG BY ANESTHESIA RESIDENT TEAM. PATIENT MOVED OUT OF OPERATING ROOM AND DOWN H HALLWAYS TO TRANSFER TO ICU. BETWEEN H AND G IT WAS NOTED THE PATIENT WAS DESATURATING AND THE VITAL SIGNS WERE BEGINNING TO DECOMPENSATE. AT THIS TIME, IT WAS NOTED THE AMBU BAG WAS NOT VENTILATING THE PATIENT. ALL CONNECTION AND OXYGEN SUPPLY WERE VERIFIED AND CHECKED. A RESIDENT RAN TO THE OPERATING ROOM TO GET A NEW AMBU BAG AND THE TEAM RUSHED THE PATIENT BACK INTO THE OPERATING ROOM. THE NEW AMBU BAG APPEARED TO FUNCTIONING PROPERLY. VENTILATION WAS VERIFIED BY AUSCULTATION AND HE WAS VENTILATED UNTIL OXYGEN SATURATIONS IMPROVED. THE PATIENT WAS THEN TRANSFERRED TO THE ICU WITHOUT ISSUE.
Patient problems: Decreased Respiratory Rate
MAUDE report 16114398 ↗MalfunctionNov 18, 2022Defective Component; Defective Component
BVM MALFUNCTIONED AND PATIENT WAS NOT ABLE TO BE VENTILATED FOR AN UNKNOWN REASON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15827984 ↗MalfunctionOct 17, 2022Defective Component; Difficult to Open or Close; Defective Component; Difficult to Open or Close
THE DUCKBILLED VALVE WOULDN'T OPEN DURING A CODE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15614750 ↗MalfunctionOct 18, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CUSTOMER STATES THAT MASKS ON RESUS BAGS ARE WARPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 7980434 ↗
Other Ventilator, Emergency, Manual (Resuscitator) devices
- Curaplex AF2000Bound Tree Medical, LLC · Ventilator, emergency, manual (resuscitator)1 recall6 reports
- Ventlab AF5140MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall5 reports
- Ventlab AF5142MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall2 reports
- Ventlab AF2140MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall2 reports
- Ventlab AF1140MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Ventlab AF2142MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 death1 report
- Ventlab VN3100OBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Medline CPRM1126FPWMMedline Industries, Inc. · Ventilator, emergency, manual (resuscitator)1 recall1 report