Skip to content

Ventlab SS3200MB-MMCRecalls & reported problems

Ventlab, LLC · INFANT SAFE SPOT W/MANOMETER, MASK,BAG RESERVOIR AND STICKER · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889483105771), received between Apr 11, 2026 and Apr 11, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 11, 2023Open, Classified

    A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

    Z-0365-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Material Integrity Problem1

Patient problems

  • Low Oxygen Saturation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2026Material Integrity Problem

    IT WAS REORTED THAT WHEN STAFF OPENED THE PACKAGING FOR PEDIATRIC BVM THE AMBU BAG MALFUNCTIONED- VALVE BROKE OFF. REPORTS PEDIATRIC PATIENT WAS IN RESPIRATORY DISTRESS. WHEN STAFF OPENED THE PACKAGING FOR PEDIATRIC BVM THE AMBU BAG MALFUNCTIONED- VALVE BROKE OFF, RESULTING IN INABILITY TO VENTILATE PATIENT. USED A LARGER PRODUCT AND PATIENT MAINTAINED PATIENT SAFETY.

    Patient problems: Low Oxygen Saturation

    MAUDE report 24851037 ↗

Other Ventilator, Emergency, Manual (Resuscitator) devices

All Ventilator, Emergency, Manual (Resuscitator) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page