Malibu 69-5540Recalls & reported problems
Seaspine Orthopedics Corporation · UNI-PLANAR DLTA SCREW,5.5 X 40 · Class 2 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889981008666), received between Feb 12, 2020 and Feb 12, 2020 — classified as 1 injury.
The most frequently reported specific device problem is “Mechanical Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Mechanical Problem1
Patient problems
- Failure of Implant1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryFeb 12, 2020Mechanical Problem; Mechanical Problem
PATIENT UNDERWENT SPINAL SURGERY USING THE MALIBU SPINAL SYSTEM WITH THE DAYTONA DEFORMITY SYSTEM. PATIENT HAD A PSEUDOARTHROSIS AT THE BOTTOM (L4/5) OF A LONG SCOLI CONSTRUCT. FUSION DID NOT OCCUR AND TWO OF THE SCREWS FAILED AND BROKE POST-OPERATIVELY. INDEX SURGERY DATE IS UNKNOWN. A REVISION SURGERY TOOK PLACE ON (B)(6) 2020, WHICH CONSISTED OF THE REMOVAL OF BI-LATERAL BROKEN SCREWS AT THE BOTTOM OF THE EXISTING SCOLI CONSTRUCT (T12). A PORTION OF THE SCREW SHAFTS WERE LEFT IN THE PEDICLE AS THE FRACTURE OCCURRED BELOW THE BONEY SURFACE. BILATERAL SCREWS WERE PLACED AT L1 / L2 / L3 (6.5 AND 7.5 MM X45 MM SCREWS). BI-LATERAL MARINER OUTRIGGER Z-RODS + BILATERAL {(1) OPEN/CLOSED 4-HOLE & (1) CLOSED/CLOSED 4-HOLE LATERAL CONNECTORS} WERE ALSO PLACED ON EACH SIDE. THE REVISION SURGERY WAS SUCCESSFUL AND THE PATIENT IS HEALING WELL, WITH NO FURTHER PLANS TO REVISE.
Patient problems: Failure of Implant; Patient Problem/Medical Problem
MAUDE report 9701452 ↗
Other Orthosis, spondylolisthesis spinal fixation devices
- CD HORIZON® Spinal System 7540020Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation257 reports
- MANTIS Redux 48289999Stryker Corporation · Orthosis, spondylolisthesis spinal fixation76 reports
- EXPEDIUM TOP NOTCH 179771555Depuy Spine, LLC · Orthosis, spondylolisthesis spinal fixation41 reports
- EXPEDIUM 179712650Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation31 reports
- EXPEDIUM VERSE 199721530Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4 5 48130000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4.5 System - XIA CT 48850000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation27 reports
- CD HORIZON® Spinal System 7576545Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation26 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.