Skip to content

Coral 18-12-6550Recalls & reported problems

Seaspine Orthopedics Corporation · Solid Polyaxial Screw, 6.50mm x 50mm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889981016012), received between May 31, 2024 and May 31, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Implant Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 31, 2024Fracture; Fracture

    ON (B)(6) 2023 A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S CORAL AND MALIBU POSTERIOR FIXATION SYSTEMS. AN MRI WAS PERFORMED AN UNKNOWN TIME LATER AFTER THE PATIENT REPORTED PAIN. THE 18-12-6550 SOLID POLYAXIAL SCREW, 6.50MM X 50MM WAS OBSERVED TO HAVE FRACTURED. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2024 AND TWO 10-17-0001 LOCKING SCREW ASSEMBLIES WERE OBSERVED TO BE LOOSE AS WELL. IT WAS NOT POSSIBLE TO REMOVE THE ENTIRE BROKEN SCREW, SO THE SURGEON LEFT THE EMBEDDED PIECE IN THE PATIENT AND INSERTED A NEW SCREW AT A DIFFERENT TRAJECTORY. THE SCREW, SCREW ASSEMBLIES, AND A 12-1040 PRECONTOURED ROD, 5.5 X 40MM WERE EXPLANTED.

    Patient problems: Implant Pain

    MAUDE report 19441571 ↗

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →

Sources and updates