Skip to content

Shoreline ACS 85-3016Recalls & reported problems

Seaspine Orthopedics Corporation · 3.5mm Variable Screw, Self Tapping, 16mm · Class 2 · Implant

All models of Shoreline ACS (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889981125172), received between Oct 6, 2023 and Apr 17, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Off-Label Use1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2025Off-Label Use; Off-Label Use

    IT WAS REPORTED ON (B)(6) 2024 THAT THREE (3) SHORELINE 3.5MM VARIABLE ANGLE (VA), SELF-TAPPING, 16MM SCREWS WERE INADVERTENTLY IMPLANTED INTO AN ADMIRAL CERVICAL PLATE DURING A MULTI-LEVEL ACDF PROCEDURE (C3-C7). THE SHORELINE SCREWS WERE USED IN COMBINATION WITH FIVE (5) ADMIRAL 4.0MM SCREWS IN A 3-LEVEL ADMIRAL PLATE CONSTRUCT. ACCORDING TO THE DISTRIBUTOR PRESENT IN THE OPERATING ROOM, THE SHORELINE SCREWS WERE PLACED PROXIMALLY, APPEARED FULLY SEATED, AND ALL LOCKING COVERS WERE VISIBLY ENGAGED. THE SURGEON ELECTED TO RETAIN THE CONSTRUCT AFTER CONFIRMING INTRAOPERATIVE STABILITY AND EXPRESSED CONFIDENCE IN THE FIXATION. NO ADVERSE PATIENT OUTCOME HAS BEEN REPORTED TO DATE. THIS EVENT INVOLVED THE OFF-LABEL USE OF SHORELINE SCREWS IN AN ADMIRAL PLATE SYSTEM, A COMBINATION THAT HAS NOT BEEN VALIDATED OR TESTED FOR CROSS-COMPATIBILITY. THE DISCREPANCY WAS IDENTIFIED POSTOPERATIVELY DURING A FOLLOW-UP DISCUSSION WITH THE DISTRIBUTOR AND TERRITORY MANAGER. NO PATIENT INJURY, EXPLANTATION, OR ADDITIONAL INTERVENTION WAS REQUIRED. AT THE TIME OF INITIAL COMPLAINT REVIEW, THE EVENT WAS DEEMED NON-REPORTABLE DUE TO THE ABSENCE OF HARM AND THE MECHANICAL INTEGRITY OF THE CONSTRUCT. HOW

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21859511 ↗
  • MalfunctionOct 6, 2023Fracture; Fracture

    ON 15 SEP 2023 SEASPINE WAS MADE AWARE THAT A SHORELINE SYSTEM SCREW WAS FRACTURED AT THE 1-YEAR POST-OPERATIVE APPOINTMENT. ADDITIONAL INFORMATION WAS RECEIVED ON 26 SEP 2023 THAT SCREWS WERE FRACTURED ON THE BOTTOM LEVELS OF BOTH IMPLANTS. THE INDEX SURGERY OCCURRED ON (B)(6) 2022. NO FURTHER INFORMATION ON THE ORIGINAL PROCEDURE WAS PROVIDED. IT WAS NOTED THAT THE FIRST SCREW WAS FRACTURED ON (B)(6) 2023. NO REPORT HAD BEEN RECEIVED BY THE COMPANY AT THAT TIME. PATIENT CONDITION WAS NOT PROVIDED. A REVISION SURGERY OCCURRED ON (B)(6) 2023 CONSISTING OF A TWO-LEVEL CORPECTOMY USING ULRICH SOLIDITY AND SEASPINE ADMIRAL PLATE, C5-T1 PCF.

    Patient problems: Insufficient Information

    MAUDE report 17889684 ↗

Other Intervertebral fusion device with integrated fixation, cervical devices

All Intervertebral fusion device with integrated fixation, cervical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page