Mariner 41-3010Recalls & reported problems

Seaspine Orthopedics Corporation · Polyaxial Head · Class 2

All models of Mariner (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889981155117), received between May 8, 2026 and May 8, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Damaged Thread” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 28, 2026Open, Classified

    Due to manufacturing error, polyaxial head may popoff from the screw intraoperatively following insertion.

    Z-2762-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Damaged Thread1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Damaged Thread

    ON 10 APR 2026 SEASPINE/ORTHOFIX WAS MADE AWARE OF A 30-MINUTE SURGICAL DELAY INVOLVING THE MARINER SPINAL SYSTEM. PER THE REPORTER, AFTER THE SURGEON IMPLANTED A SCREW DURING AN L3 PELVIS CASE ON (B)(6) 2026, HE EXPERIENCED ISSUES THREADING THE SET SCREW ONTO THE TULIP. THE SURGEON ATTEMPTED TO RESOLVE THE ISSUE USING TWO DIFFERENT SET SCREWS, HOWEVER, NEITHER SET SCREW WOULD THREAD ONTO THE TULIP. THE SURGEON PROCEEDED TO REMOVE THE SCREW AND TULIP AND REPLACED IT WITH A NEW ONE. THE CASE WAS ABLE TO BE COMPLETED SUCCESSFULLY AND NO PATIENT INJURIES WERE REPORTED. THE ISSUE RESULTED IN A DELAY OF APPROXIMATELY 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25129353 ↗