Skip to content

Mariner recallZ-2762-2026 · Seaspine Orthopedics Corporation · initiated May 28, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Seaspine Orthopedics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to manufacturing error, polyaxial head may popoff from the screw intraoperatively following insertion.

FDA root cause category: Process change control

What the company is doing

On June 16, 2026, Orthofix issued a updated Urgent Medical Device Recall notification to provide additional information concerning the recall. Initial notification was issued on May 28, 206. Orthofix ask consignees to take the following actions: 1. Discontinue use of affected lots, 2.Immediately identify and quarantine any affected product. 3. If product is identified, Orthofix will facilitate the return. 4. Complete and return the attached Acknowledgement and Receipt Form. 5. This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred.

Recall details

Recall number
Z-2762-2026
Status
Open, Classified
Initiated by the company
May 28, 2026
Posted by the FDA
Jul 20, 2026
Recalling firm
Seaspine Orthopedics Corporation
Quantity
796 units
Product code
NKB
FDA submissions
K173882
Distribution
Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, California, Colorado, Connecticut, Florida, Illinois, Indiana, Kansas, Maryland, Michigan, Missouri, Mississippi, Montana, Nevada, Ohio, Oklahoma, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, and Wyoming. The countries of Chile and Australia.

Product description

SeaSpine Polyaxial Head REF Number: 41-3010

Code information (models, lots, serial numbers)

UDI: 10889981155117; Affected Lots: AW199682G, AW199684G.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K173882 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page