Skip to content

NorthStar PC1-105534Recalls & reported problems

Seaspine Orthopedics Corporation · Poly. Screw, Favored Angle 5.5mm x 34mm · Class 2 · Implant

All models of NorthStar (24)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889981197308), received between Jun 21, 2024 and Jun 21, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 21, 2024Fracture; Fracture

    ON 23 MAY 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S NORTHSTAR POSTERIOR CERVICAL FUSION SYSTEM. A FAVORED ANGLE SCREW BROKE JUST BELOW THE SCREW HEAD DURING INSERTION INTO THE LEFT SIDE OF THE T1 VERTEBRAE. A SECOND FAVORED ANGLE SCREW BROKE AT THE SAME POINT DURING INSERTION INTO THE RIGHT SIDE OF THE T1 VERTEBRAE. BOTH SCREW SHAFTS WERE REMOVED FROM THE PATIENT. THIS RESULTED IN A 20 TO 30-MINUTE SURGICAL DELAY. A STRAIGHT PROBE HAD BEEN USED TO CREATE BOTH SCREW PILOT HOLES. NO PATIENT INJURY WAS REPORTED. TWO SCREWS WERE RETURNED FOR EVALUATION; IT IS UNKNOWN WHICH SCREW IS ASSOCIATED WITH WHICH FRACTURE. THIS REPORT IS FOR LOT CR46588D.

    Patient problems: Insufficient Information

    MAUDE report 19584664 ↗
  • MalfunctionJun 21, 2024Fracture; Fracture

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S NORTHSTAR POSTERIOR CERVICAL FUSION SYSTEM. A POLYAXIAL, FAVORED ANGLE SCREW BROKE JUST BELOW THE SCREW HEAD DURING INSERTION INTO THE LEFT SIDE OF THE T1 VERTEBRAE. A SECOND POLYAXIAL, FAVORED ANGLE SCREW BROKE AT THE SAME POINT DURING INSERTION INTO THE RIGHT SIDE OF THE T1 VERTEBRAE. BOTH SCREW SHAFTS WERE REMOVED FROM THE PATIENT. THIS RESULTED IN A 20 TO 30-MINUTE SURGICAL DELAY. A STRAIGHT PROBE HAD BEEN USED TO CREATE BOTH SCREW PILOT HOLES. NO PATIENT INJURY WAS REPORTED. TWO SCREWS WERE RETURNED FOR EVALUATION; IT IS UNKNOWN WHICH SCREW IS ASSOCIATED WITH WHICH FRACTURE. THIS REPORT IS FOR LOT CR46075C.

    Patient problems: Insufficient Information

    MAUDE report 19584604 ↗

Other Posterior cervical screw system devices

All Posterior cervical screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page