Admiral AP2-100506Recalls & reported problems
Seaspine Orthopedics Corporation · Temporary Fixation Pin · Class 1
All models of Admiral (19)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10889981277109), received between Jul 21, 2023 and Jan 13, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Material Fragmentation2
Patient problems
- Foreign Body In Patient2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2026Material Fragmentation
ON (B)(6) 2025, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S ADMIRAL ACP SYSTEM. IT WAS REPORTED THAT TWO TEMPORARY FIXATION PINS (AP2-100506) FRACTURED AND THE TIPS WERE LEFT INSIDE THE PATIENT'S BONE. NO PATIENT INJURY WAS REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 24059565 ↗MalfunctionNov 18, 2025Fracture; Fracture
ON (B)(6) 2025, A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S ADMIRAL ACP SYSTEM. DURING THE PROCEDURE A TEMPORARY FIXATION PIN (AP2100506) WAS PLACED INTO THE LOWER SCREW HOLE OF THE PLATE. SCREWS WERE PLACED INTO ALL OTHER SCREW HOLES. UPON REMOVAL OF THE TEMPORARY FIXATION PIN IT WAS OBSERVED THAT A SIGNIFICANT PORTION OF THE PIN HAD BROKEN AND REMAINED IN BONE. THE REPORTER NOTED THAT THE BROKEN FRAGMENT WAS NOT RETRIEVED FROM THE SURGICAL SITE. A NEW TRAJECTORY WAS DRILLED AND A SHORTENED SCREW WAS PLACED. THE REPORTER NOTED THAT NO OBSERVABLE INJURY RESULTED FROM THE BROKEN FIXATION PIN, HOWEVER THE BROKEN FRAGMENT REMAINS IN THE BONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23583989 ↗MalfunctionMar 7, 2025Material Fragmentation; Material Fragmentation
CUSTOMER REPORTS THAT DURING A SURGERY THAT TOOK PLACE ON (B)(6) 2025, AN ADMIRAL TEMPORARY FIXATION PIN BROKE WITH THE DISTAL PORTION OF THE PIN REMAINING IN THE BONE. THERE WAS NO WAY TO RETRIEVE THE PIECE THAT REMAINED IN THE BONE; HOWEVER, THE SURGEON DID NOT FEEL THERE WAS ANY RISK OF THAT PORTION BACKING OUT OF THE BONE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN SURGERY AND NO PATIENT INJURY REPORTED. THE ADMIRAL TEMPORARY FIXATION PIN WAS NOT SAVED BY THE CUSTOMER DUE TO THE SIZE AND SHARPNESS, THEREFORE, THE PRODUCT IS NOT AVAILABLE FOR EVALUATION. ADDITIONALLY, THE CUSTOMER STATES THE LOT NUMBER WAS NOT LEGIBLE AS THE LASER MARKING WAS NO LONGER DARK AND WAS INCREDIBLY SMALL.
Patient problems: Foreign Body In Patient
MAUDE report 21554470 ↗MalfunctionJul 21, 2023Fracture; Fracture
ON (B)(6) 2023, A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S ADMIRAL ACP SYSTEM. DURING THE PROCEDURE, AN AP2-100506 TEMPORARY FIXATION PIN WAS INSERTED THROUGH A SCREW HOLE; THE TEMPORARY FIXATION PIN WAS THEN REMOVED WITHOUT BEING AWARE THAT THE PIN HAD FRACTURED AND A FRAGMENT WAS LEFT IN THE SURGICAL SITE. THE SURGEON CONTINUED THROUGH THE PROCEDURE BY INSERTING AN AP2-100314 ADMIRAL ACP DRILL, 14MM INTO THE SAME SCREW HOLE. THE ACP DRILL MADE CONTACT WITH THE TEMPORARY FIXATION PIN AND CREATED ENOUGH FRICTION TO CAUSE IT TO FRACTURE AS WELL. FRAGMENTS OF BOTH THE AP2-100506 TEMPORARY FIXATION PIN AND THE AP2-100314 ACP DRILL REMAIN IN-SITU. NO REPORTS OF TISSUE DAMAGE OR PATIENT INJURY WAS REPORTED AND THERE ARE NO PLANS FOR REVISION OR REMOVAL AT THIS TIME. SEASPINE WAS MADE AWARE OF THIS INCIDENT ON (B)(6) 2023. REPORT # AP2-100506 TEMPORARY FIXATION PIN. REPORT #3012120772-2023-00014 AP2-100314 ACP DRILL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17372603 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page