Skip to content

Meridian RA2-006014Recalls & reported problems

Seaspine Orthopedics Corporation · Modular Plate Compressor, Short · Class 1

All models of Meridian (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889981284503), received between Jun 4, 2026 and Jun 4, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Mechanical Problem

    ON (B)(6) 2026, DURING A SPINAL FUSION PROCEDURE, THE REPORTER EXPERIENCED DIFFICULTY USING A MODULAR PLATE COMPRESSOR TO SECURE A PLATE TO AN IMPLANT. MULTIPLE ATTEMPTS WERE MADE USING ADDITIONAL IMPLANTS AND PLATES; HOWEVER, THE CONDITION PERSISTED. AN ALTERNATE INSTRUMENT FROM A DIFFERENT INSTRUMENT SET WAS OBTAINED AND SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. THE EVENT REPORTEDLY RESULTED IN AN APPROXIMATE 30-MINUTE SURGICAL DELAY, PROMPTING THIS REPORT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO REPORTED PATIENT INJURY OR ADVERSE CLINICAL OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25394810 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page