Skip to content

Reef TH TA2-005002Recalls & reported problems

Seaspine Orthopedics Corporation · TA Removal Tool · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889981318406), received between Mar 26, 2025 and Sep 10, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility1
  • Separation Failure1

Patient problems

  • Dural Tear1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 10, 2026Separation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722602 ↗
  • InjuryMar 26, 2025Device-Device Incompatibility; Device-Device Incompatibility

    DURING A SURGERY THAT TOOK PLACE ON (B)(6) 2025, THE SURGEON WAS ATTEMPTING TO INSERT A TRIAL INTO THE PATIENT USING THE REEF TA INSERTER AND UPON MALLETING THE TRIAL INTO THE DISC SPACE, IT UNEXPECTEDLY DISLODGED AND MOVED INTO THE DURA. THE SURGEON WAS ABLE TO RETRIEVE THE TRIAL USING CURETTES AND OTHER INSTRUMENTS, BUT THE TA REMOVAL TOOL WAS NOT EFFECTIVE AS IT COULD NOT SECURELY GRIP THE BACK OF THE TRIAL. AFTER THE DURAL TEAR, THE SURGEON ATTEMPTED TO LOAD ANOTHER TRIAL, HOWEVER, THE SURGEON WAS ABLE TO DETACH IT BY HAND, INDICATING A POTENTIAL ISSUE WITH THE LOCKING MECHANISM. THIS ISSUE CAUSED A ONE HOUR DELAY WHILE THE SURGEON HAD TO REPAIR THE DURA. CUSTOMER CONFIRMS THE SURGEON FOLLOWED THE APPROPRIATE SURGICAL TECHNIQUE FOR LOCKING THE TRIAL ON THE INSERTER AND THAT THE PATIENT IS DOING WELL AFTER THE SURGERY.

    Patient problems: Dural Tear

    MAUDE report 21702148 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page