Skip to content

Dextera MicroCutter 5/80 Stapler FG-025302Recalls & reported problems

Aesdex, LLC · Endoscopic Surgical Stapler · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 11814900003025), received between Jan 5, 2017 and May 19, 2017 — classified as 1 death, 1 injury.

The most frequently reported specific device problem is “Positioning Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Positioning Problem1

Patient problems

  • Blood Loss1
  • Renal Failure1
  • Respiratory Failure1
  • Sepsis1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 19, 2017Adverse Event Without Identified Device or Use Problem; Positioning Problem; Adverse Event Without Identified Device or Use Problem; Positioning Problem

    THE PATIENT WAS UNDERGOING A RIGHT LUNG MIDDLE LOBECTOMY BY ROBOTIC VIDEO-ASSISTED THORACIC SURGERY. THE MICROCUTTER 5/80 STAPLER WAS ADVANCED THROUGH THE ACCESSORY PORT TROCAR TO A BRANCH OF THE SUPERIOR PULMONARY VEIN FOLLOWING DISSECTION OF TISSUE AROUND THE TARGET VEIN. WHILE POSITIONING THE STAPLER AT THE SELECTED LOCATION FOR TRANSECTION OF THE VEIN AND PRIOR TO FIRING THE STAPLER, BLEEDING OCCURS WHICH APPEARS TO HAVE ORIGINATED FROM BEHIND THE TARGET VEIN. IT APPEARS THAT A SMALL BRANCH VESSEL, LOCATED BEHIND THE TARGET VEIN, MAY HAVE AVULSED WHEN THE MICROCUTTER 5/80 STAPLER WAS BEING POSITIONED, RESULTING IN BLEEDING. THE STAPLER IS SUBSEQUENTLY FIRED RESULTING IN A SUCCESSFUL STAPLE LINE. THE SURGEON THEN USES PRESSURE TO CONTROL THE BLEEDING. ADDITIONAL BLEEDING IS NOTED AT THE DISTAL END OF THE TRANSECTED VESSEL. THE SURGEON CONTROLS THE BLEEDING THROUGH THE USE OF DIRECT PRESSURE, CAUTERY, TOPICAL HEMOSTATIC AGENTS AND CLIPS. THE SURGERY WAS COMPLETED AND IT WAS REPORTED THAT THE PATIENT DID NOT EXPERIENCE ANY FURTHER COMPLICATIONS AS A RESULT OF THE BLEEDING AND IS DOING WELL.

    Patient problems: Blood Loss

    MAUDE report 6580091 ↗
  • DeathJan 5, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE SURGEON REPORTED AN ADVERSE EVENT THAT OCCURRED DURING A RIGHT UPPER LOBECTOMY AND LOWER LOBE WEDGE RESECTION VIA A VIDEO-ASSISTED THORACOSCOPIC SURGERY (VATS). WHILE POSITIONING THE STAPLER BEHIND THE RIGHT UPPER LOBE SEGMENTAL PULMONARY ARTERY BRANCH, THE STAPLER WAS PUSHED SOMEWHAT FORCEFULLY AND THE VESSEL TORE DISTAL TO THE SITE TARGETED FOR TRANSECTION OF THE VESSEL. THE STAPLES WERE THEN DEPLOYED IN THE PROXIMAL PORTION OF THE VESSEL, RESULTING IN A SUCCESSFUL STAPLE LINE. THE SURGICAL PROCEDURE WAS CONVERTED TO AN OPEN THORACOTOMY DUE TO THE DISTAL TEAR AND SUBSEQUENT BLEEDING. FOLLOWING THE CONVERSION, THE SURGEON OBSERVED THAT THE STAPLES REMAINED INTACT AND THERE WAS NO ISSUE WITH THE DISTAL STAPLE LINE, HOWEVER THE TEAR WAS NOTED DISTAL TO THE STAPLE LINE. IN A SUBSEQUENT COMMUNICATION ON (B)(6) 2017, THE SURGEON REPORTED THAT THE PATIENT HAD EXPIRED. IT WAS REPORTED BY THE HOSPITAL PERSONNEL THAT THE PATIENT WAS EXPIRED FROM SEPSIS ON (B)(6) 2016.

    Patient problems: Renal Failure; Respiratory Failure; Sepsis; Unspecified Infection

    MAUDE report 6227739 ↗

FDA clearance and classification

The FDA regulates product code GAG (stapler, surgical) under 21 CFR 878.4740 as a Class II device. 21 CFR 878.4740 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: A surgical stapler for internal use is a specialized prescription device used to deliver compatible staples to internal tissues during surgery for resection, transection, and creating anastomoses.

Common questions about the Dextera MicroCutter 5/80 Stapler FG-025302

Has the Dextera MicroCutter 5/80 Stapler FG-025302 been recalled?

No FDA recall record in our data names this exact device identifier (DI 11814900003025), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Dextera MicroCutter 5/80 Stapler FG-025302?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jan 5, 2017 and May 19, 2017. In those reports, the most common specific device problem is “Positioning Problem” (1); the most common patient problems are “Blood Loss” (1), “Renal Failure” (1) and “Respiratory Failure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dextera MicroCutter 5/80 Stapler FG-025302?

1 report is classified as a death, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Dextera MicroCutter 5/80 Stapler FG-025302, and is it still sold?

The FDA's device database lists Aesdex, LLC as the company responsible for it (brand Dextera MicroCutter 5/80 Stapler, model FG-025302). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Stapler, Surgical devices

All Stapler, Surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page