RUSCH recallZ-1011-2023 · Teleflex LLC · initiated Dec 8, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Teleflex LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for pilot balloon non-inflation or cuff non-deflation.

FDA root cause category: Process control

What the company is doing

Teleflex issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/08/2022 by FedEx 2nd day mail. The notices explain the issues, risk, and requested the following: "Action list number 1 Medical facilities 1. We request that you immediately check your inventory for product within the scope of this FSCA. Users should cease use and distribution of affected product and immediately quarantine the affected product. 2. If you have impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and fax it to 1-855-419-8507, including Attn: Customer Service or using the email address [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you do not have impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and return the form to Teleflex at the contact details provided. 4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product. Action list number 2 Distributors 1. Provide this field safety notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you. 2. We request that you immediately check your inventory for impacted product. Cease use and distribution of impacted product and immediately quarantine the affected product. You may then return all product in scope. 3. As a distributor, you are then required to confirm to Teleflex that you have completed the field activity outlined in actions 1 and 2 of this Action List Number 2. Upon completion of your actions, please forward the completed Acknowledgement Form to the e-mail address below. Important - Please ensure you only list batch numbers in scope of this Field Safety Notice

Recall details

Recall number
Z-1011-2023
Status
Open, Classified
Initiated by the company
Dec 8, 2022
Posted by the FDA
Jan 23, 2023
Recalling firm
Teleflex LLC
Quantity
39,690 units
Product code
BTR
FDA submissions
K961837
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WA, WI, WV and PR. The countries of Albania, Armenia, Austria, Azerbaijan, Bahamas, Barbados, Belgium, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Croatia, Cyprus, Czech, Denmark, El Salvador, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, Ireland, Italy Israel, Jordan, Kuwait, Lebanon, Liechtenstein, Malaysia, Mayotte, Mexico, Moldova, Morocco, Nepal, Netherlands, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russia, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, South Africa, South Korea, Sweden, Switzerland, Taiwan Thailand, Turkey, Ukraine, United Arab Emirates.

Product description

Endotracheal Tube labeled as the following: a. Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal; b. Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff; c. RUSCHELIT Super Safety Clear Tracheal Tube, oral/nasal, Murphy. Intended for patients requiring endotracheal intubation

Code information (models, lots, serial numbers)

Show the full code information (6k characters)

a. Product Code, Batch Number (UDI): 504550, KME21D1964 ((01)4026704547579(17)260328(10)KME21D1964), 504560, KME21E1412 ((01)4026704547593(17)260428(10)KME21E1412), 504580, KME21D1784 ((01)4026704547630(17)260328(10)KME21D1784), 504580, KME21E1286 ((01)4026704547630(17)260428(10)KME21E1286), 504585, KME21E1295 ((01)4026704547647(17)260428(10)KME21E1295), 504590, KME21E1114 ((01)4026704547654(17)260428(10)KME21E1114); b. Product Code, Batch Number (UDI): 112082050, KME21F0727 ((01)14026704341389(17)260528(10)KME21F0727), 112082050, KME21F0321 ((01)14026704341389(17)260528(10)KME21F0321), 112082055, KME21E0864 ((01)14026704341396(17)260428(10)KME21E0864), 112082055, KME21F0946 ((01)14026704341396(17)260528(10)KME21F0946), 112082055, KME21E1230 ((01)14026704341396(17)260428(10)KME21E1230), 112082060, KME21E0660 ((01)14026704341402(17)260428(10)KME21E0660), 112082060, KME21E1391 ((01)14026704341402(17)260428(10)KME21E1391), 112082065, KME21E0607 ((01)14026704341419(17)260428(10)KME21E0607), 112082065, KME21E0685 ((01)14026704341419(17)260428(10)KME21E0685), 112082065, KME21F0384 ((01)14026704341419(17)260528(10)KME21F0384), 112082065, KME21F0516 ((01)14026704341419(17)260528(10)KME21F0516), 112082065, KME21F0694 ((01)14026704341419(17)260528(10)KME21F0694), 112082065, KME21F0834 ((01)14026704341419(17)260528(10)KME21F0834), 112082065, KME21F0930 ((01)14026704341419(17)260528(10)KME21F0930), 112082070, KME21D2935 ((01)14026704341426(17)260328(10)KME21D2935), 112082070, KME21E0002 ((01)14026704341426(17)260428(10)KME21E0002), 112082070, KME21E0363 ((01)14026704341426(17)260428(10)KME21E0363), 112082070, KME21E1390 ((01)14026704341426(17)260428(10)KME21E1390), 112082070, KME21D1580 ((01)14026704341426(17)260328(10)KME21D1580), 112082070, KME21D1652 ((01)14026704341426(17)260328(10)KME21D1652), 112082070, KME21D1706 ((01)14026704341426(17)260328(10)KME21D1706), 112082070, KME21D1878 ((01)14026704341426(17)260328(10)KME21D1878), 112082070, KME21D2020 ((01)14026704341426(17)260328(10)KME21D2020), 112082070, KME21D2929 ((01)14026704341426(17)260328(10)KME21D2929), 112082070, KME21E0001 ((01)14026704341426(17)260428(10)KME21E0001), 112082070, KME21F0102 ((01)14026704341426(17)260528(10)KME21F0102), 112082070, KME21F0140 ((01)14026704341426(17)260528(10)KME21F0140), 112082070, KME21F0239 ((01)14026704341426(17)260528(10)KME21F0239), 112082070, KME21F0434 ((01)14026704341426(17)260528(10)KME21F0434), 112082075, KME21E0288 ((01)14026704341433(17)260428(10)KME21E0288), 112082075, KME21E1150 ((01)14026704341433(17)260428(10)KME21E1150), 112082075, KME21E1178 ((01)14026704341433(17)260428(10)KME21E1178), 112082075, KME21E1186 ((01)14026704341433(17)260428(10)KME21E1186), 112082075, KME21F0557 ((01)14026704341433(17)260528(10)KME21F0557), 112082075, KME21D1705 ((01)14026704341433(17)260328(10)KME21D1705), 112082075, KME21D1780 ((01)14026704341433(17)260328(10)KME21D1780), 112082075, KME21D1923 ((01)14026704341433(17)260328(10)KME21D1923), 112082075, KME21D1956 ((01)14026704341433(17)260328(10)KME21D1956), 112082075, KME21D2014 ((01)14026704341433(17)260328(10)KME21D2014), 112082075, KME21D2018 ((01)14026704341433(17)260328(10)KME21D2018), 112082075, KME21D2262 ((01)14026704341433(17)260328(10)KME21D2262), 112082075, KME21E0021 ((01)14026704341433(17)260428(10)KME21E0021), 112082075, KME21E0498 ((01)14026704341433(17)260428(10)KME21E0498), 112082075, KME21E0469 ((01)14026704341433(17)260428(10)KME21E0469), 112082075, KME21F0351 ((01)14026704341433(17)260528(10)KME21F0351), 112082075, KME21E1264 ((01)14026704341433(17)260428(10)KME21E1264), 112082075, KME21F0184 ((01)14026704341433(17)260528(10)KME21F0184), 112082075, KME21F0385 ((01)14026704341433(17)260528(10)KME21F0385), 112082075, KME21F0594 ((01)14026704341433(17)260528(10)KME21F0594), 112082075, KME21F0725 ((01)14026704341433(17)260528(10)KME21F0725), 112082080, KME21D2012 ((01)14026704341440(17)260328(10)KME21D2012), 112082080, KME21F0981 ((01)14026704341440(17)260528(10)KME21F0981), 112082080, KME21D1653 ((01)14026704341440(17)260328(10)KME21D1653), 112082080, KME21D1946 ((01)14026704341440(17)260328(10)KME21D1946), 112082080, KME21D2011 ((01)14026704341440(17)260328(10)KME21D2011), 112082080, KME21E0584 ((01)14026704341440(17)260428(10)KME21E0584), 112082080, KME21E0687 ((01)14026704341440(17)260428(10)KME21E0687), 112082080, KME21E1452 ((01)14026704341440(17)260428(10)KME21E1452), 112082080, KME21E0770 ((01)14026704341440(17)260428(10)KME21E0770), 112082080, KME21F1006 ((01)14026704341440(17)260528(10)KME21F1006), 112082085, KME21E0902 ((01)14026704341457(17)260428(10)KME21E0902), 112082085, KME21E1291 ((01)14026704341457(17)260428(10)KME21E1291), 112082085, KME21F0741 ((01)14026704341457(17)260528(10)KME21F0741), 112082095, KME21E1393 ((01)14026704341471(17)260428(10)KME21E1393); c. Product Code, Batch Number (UDI): 112482065LA, KME21E0467 ((01)14026704629142(17)260428(10)KME21E0467), 112482070LA, KME21E0450 ((01)14026704629159(17)260428(10)KME21E0450), 112482075LA, KME21E0372 ((01)14026704629166(17)260428(10)KME21E0372), 112482075LA, KME21E0378 ((01)14026704629166(17)260428(10)KME21E0378), 112482075LA, KME21E0451 ((01)14026704629166(17)260428(10)KME21E0451), 112482075LA, KME21E0489 ((01)14026704629166(17)260428(10)KME21E0489), 112482080LA, KME21E0381 ((01)14026704629173(17)260428(10)KME21E0381), 112482085LA, KME21F0101 ((01)14026704629180(17)260528(10)KME21F0101), 112482085LA, KME21E0779 ((01)14026704629180(17)260428(10)KME21E0779),112482085LA, KME21E0956 ((01)14026704629180(17)260428(10)KME21E0956), 112482085LA, KME21F0520 ((01)14026704629180(17)260528(10)KME21F0520), 112482085LA, KME21F0637 ((01)14026704629180(17)260528(10)KME21F0637), 112482095LA, KME21E0742 ((01)14026704629203(17)260428(10)KME21E0742), 112482100LA, KME21D1667 ((01)14026704629210(17)260328(10)KME21D1667)

Devices named in this recall

14 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

8 devices cleared under K961837 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 91301).