Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 14953170247573), received between Dec 13, 2020 and Jul 1, 2024 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Detachment of Device or Device Component1
- Fitting Problem1
- Improper or Incorrect Procedure or Method1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJul 1, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THERAPEUTIC MUCOSECTOMY/POLYPECTOMY THE INTERNAL RUBBER FALLS OFF EASILY DURING PROCEDURES. WHEN IT IS IN USE, THE DOCTOR HAS TO INSERT AND REMOVE NEEDLES, CLAMPS, SNARES, ETC, FROM THE WORKING CHANNEL MANY TIMES. THIS FRICTION ENDS UP DETACHING THE RUBBER AND PUSHING IT THROUGH THE WORKING CHANNEL. IN ONE CASE, IT FELL INTO THE PATIENT'S COLON AND THE DOCTOR HAD TO "HUNT" IT DOWN AND REMOVE IT WITH FORCEPS. THERE WAS NO REPORTED INJURY OR HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19650325 ↗MalfunctionMar 27, 2023Physical Resistance/Sticking; Physical Resistance/Sticking
AN OLYMPUS REPRESENTATIVE REPORTED TO OLYMPUS ON BEHALF OF THE CUSTOMER THAT DURING REPROCESSING, THIS SINGLE USE BIOPSY VALVE, GOT STUCK IN BIOPSY CHANNEL. IT WAS IDENTIFIED THAT THE STAFF WAS REUSING THE VALVES WHICH IS AFFECTING THE INTEGRITY OF THE DEVICE. THE STAFF, DOCTORS, BIOMEDICAL AND WARD MANAGER WERE CONSIDERING TO START USING REUSABLE VALVES. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16624414 ↗MalfunctionOct 11, 2022Fitting Problem; Fitting Problem
THE CUSTOMER REPORTED DISSATISFACTION WITH THE FUNCTIONALITY OF THE SUBJECT DEVICE (BIOPSY VALVE). IT WAS REPORTED THE VALVE WOULD NOT FIT THE LUER LOCK SYRINGE AND WHEN USING IT WITH A POINTED SYRINGE THE WATER WOULD COME OUT FROM ALL SIDES WHICH MEANS THAT WATER CANNOT BE IRRIGATED THROUGH THE CHANNEL, PREVENTING THE CORRECT CLEANING OF THE MUCOSA TO CONTINUE THE PROCEDURE. THE REPORTED ISSUE OCCURRED DURING AN UNSPECIFIED DIAGNOSTIC PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING THE SUBJECT DEVICE. THERE WAS NO PATIENT OR USER INJURY REPORTED DUE TO THE EVENT. THIS REPORT IS BEING SUBMITTED DUE TO WATER COMING OUT OF THE BIOPSY VALVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15583893 ↗MalfunctionMar 2, 2022Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THE USER FACILITY REQUESTED A REPROCESSING IN-SERVICE FOR THE SINGLE USE BIOPSY VALVE (STERILE). AN OLYMPUS ENDOSCOPY SUPPORT SPECIALIST (ESS) VISITED THE CUSTOMER AND NOTED DURING THE IN-SERVICE, THE CUSTOMER WAS NOT DISCARDING SINGLE USE BIOPSY BUT INSTEAD REPROCESSING THEM. IT WAS REPORTED THAT THE CUSTOMER WAS NOT USING MH-948 ENDOSCOPE AW CHANNEL CLEANING ADAPTER FOR PRECLEANING OF EQUIPMENT. THERE WAS NO PATIENT INVOLVEMENT, NO HARM OR USER INJURY REPORTED DUE TO THE EVENT. ADDITIONAL INFORMATION FROM THE USER FACILITY CONFIRMED THAT THE BIOPSY CHANNEL COVERS WERE CLEANED AND STERILIZED FOR REPROCESS TRAINING PURPOSES AND THE BIOPSY CHANNEL COVERS WERE NOT REUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13653764 ↗MalfunctionDec 13, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED FROM THE USER THAT THE SUBJECT DEVICE OPENED EASILY DURING AN UNSPECIFIED PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10999771 ↗
Other Colonoscope And Accessories, Flexible/Rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page