Medfusion recallZ-0079-2024 · Smiths Medical ASD Inc. · initiated Aug 16, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (including Force Sensor BGND Test, Force Sensor Bridge Test, or Force Sensor Test). However, if the calibration shift is not large enough to trigger a System Failure Alarm, the threshold to detect an occlusion may increase, increasing the time to occlusion detection, or the threshold may decrease, leading to false occlusion alarms. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly. Out of an abundance of caution, we are notifying all customers of this potential issue.
FDA root cause category: Under Investigation by firm
What the company is doing
Smiths Medical sent an URGENT MEDICAL DEVICE CORRECTION notice to its direct consignees on 08/16/2023 by traceable mailing. The notice explained the issue, potential risk to the patient, and requested the following actions be taken: Customer Required Actions 1. Locate all Medfusion Model 3500 and Medfusion Model 4000 syringe pumps in your possession and ensure all users or potential users of these devices are immediately made aware of this notification and proposed mitigations. 2. Ensure all Medfusion Model 3500 and Medfusion Model 4000 syringe pumps in your possession have undergone all tests in the Annual Maintenance List including the new Force Sensor Check within the past 12 months, including new pumps that have not yet undergone annual maintenance. 3. Complete and return the attached Response Form to [email protected] within ten days of receipt to acknowledge your understanding of this notification. 4. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]
Recall details
- Recall number
- Z-0079-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 16, 2023
- Posted by the FDA
- Oct 6, 2023
- Recalling firm
- Smiths Medical ASD Inc.
- Quantity
- 67507 pumps
- Product code
- FRN
- FDA submissions
- K111386
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
Product description
Medfusion Syringe Pump, Model 4000-XXXX-XX: a) 0100-50, b) 0100-249, c) 0101-50, d) 0101-51, e) 0101-78, f) 0101-249, g) 0105-51, h) 0105-78, i) 0105-249, j) 0106-00, k) 0106-01, l) 0106-231, m) 0107-01, software versions: v3, v4, v5, and v6
Code information (models, lots, serial numbers)
a) 0100-50, UDI/DI 10610586035326; b) 0100-249, UDI/DI 10610586033124; c) 0101-50, UDI/DI 10610586040054; d) 0101-51, UDI/DI 10610586040818; e) 0101-78, UDI/DI 10610586040788; f) 0101-249, UDI/DI 10610586040078; g) 0105-51, UDI/DI 10610586043567; h) 0105-78, UDI/DI 10610586043574; i) 0105-249, UDI/DI 10610586043581; j) 0106-00, UDI/DI 15019517070750; k) 0106-01, UDI/DI 15019517069624; l) 0106-231, UDI/DI 15019517070996; m) 0107-01, UDI/DI 15019517154313; All serial numbers
Devices named in this recall
8 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medfusion 4000-0106-01Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls4,390 reports
- Medfusion 4000-0105-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls2,882 reports
- Medfusion 4000-0101-51Icu Medical, Inc. · Pump, infusion2 recalls1 death1,795 reports
- MEDFUSION 4000-0106-00Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls10 reports
- MEDFUSION 4000-0106-231Icu Medical, Inc. · Pump, infusion2 recalls5 reports
- Medfusion 4000-0105-249Icu Medical, Inc. · Pump, infusion2 recalls1 report
- Medfusion 4000-0105-78Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls0 reports
- MEDFUSION 4000-0107-01Icu Medical, Inc. · Pump, infusion1 recall3 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K111386 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Medfusion 3500-0600-01Icu Medical, Inc. · Pump, infusion2 recalls2,041 reports
- Medfusion 3500-0600-82Icu Medical, Inc. · Pump, infusion2 recalls89 reports
- Medfusion 3500-0600-00Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls9 reports
- Medfusion 3500-0600-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls2 reports
- PharmGuard 22-4075-0301-51Icu Medical, Inc. · Pump, infusion3 reports
- Medfusion 67-2477-0600-01Icu Medical, Inc. · Pump, infusion1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 92956).
- Z-0076-2024 · Open, ClassifiedMedfusion Syringe Pump, Model 3500-500, v3, v4, v5, and v6
- Z-0077-2024 · Open, ClassifiedMedfusion Syringe Pump, Model 3500VX-500, v3, v4, v5, and v6
- Z-0078-2024 · Open, ClassifiedMedfusion Syringe Pump, Model 3500-0600-XX: a) 0600-00; b) 0600-01; c) 0600-50; d) 0600-51; e) 0600-82; f) 0600-249; software versions: v3, v4, v5, and v6
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page