2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15019517101034), received between May 7, 2026 and May 7, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Infuse” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

4 other recalls cover devices cleared under the same FDA submission (K170982); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 10, 2025Open, Classified

    There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.

    Z-1681-2025 · Root cause: Under Investigation by firm · FDA record ↗

  2. Feb 27, 2024Open, Classified

    The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.

    Z-2420-2024 · Root cause: Software Manufacturing/Software Deployment · FDA record ↗

Most reported problems

Device problems

  • Failure to Infuse1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2026Failure to Infuse

    IT WAS FOUND DURING INVESTIGATION THAT THE REPORTED CASSETTE ATTACHMENT ERROR WAS CONFIRMED THROUGH LOGS. THERE WAS NO PATIENT INVOLVEMENT, AND NO HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25111752 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170982. One submission often covers many models, so check the product and lot details in each record.

  • Z-1662-2025 · Apr 10, 2025 · Open, Classified
    CADD Solis HSPCA Pump ***Updated 5/27/26*** Scope updated to apply specifically to customers using the wireless system. Affected products associated with the change in scope: PUM
  • Z-1667-2025 · Apr 10, 2025 · Open, Classified
    CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SO
  • Z-0466-2025 · Jun 26, 2024 · Open, Classified
    smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps
  • Z-0963-2023 · Dec 9, 2022 · Open, Classified
    CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4) REF 21-7321-24, 5) REF 21-7322-24, 6) REF 21-7323-24, 7) REF 21-7324-24, 8)

Other Pump, infusion, pca devices

All Pump, infusion, pca devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.