Skip to content

CADD recallZ-1681-2025 · Smiths Medical ASD, Inc. · initiated Apr 10, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION dated 4/10/25 was sent to customers. Actions for Users: Inform all affected CADD-Solis and CADD-Solis VIP users of this notice. Provide the instructions below: 1. Ensure all users or potential users of these products are immediately made aware of this notification. 2. Upon receipt of this notification and ongoing during regular maintenance, visually inspect the external condition of the battery pack or Wireless Communication Module and look for evidence of any physical damage. Handle a damaged device with caution. As stated in the battery pack Instructions for Use, if the battery pack housing is cracked or otherwise damaged, replace the battery pack. NEVER use a battery pack that appears damaged. A rechargeable battery pack must be replaced with either another CADD-Solis rechargeable battery pack or with 4 AA batteries. 3. Visually inspect the condition of the battery compartment for evidence of any physical damage. Contact Technical Support for repairs. 4. Always check the battery compartment for fluid or debris before inserting the batteries, and do not allow any fluid or debris to fall into the battery compartment. 5. Complete the electronic Response Form via the link provided within this email notification within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product. 6. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward your customer list to Sedgwick at [email protected]. Sedgwick will notify customers with email addresses via email and will send physical notifications to customers without email addresses in order for them to complete the Response Form Follow up Actions by Smiths Medical: Smiths Medical is continuing to investigate this matter to determine if additional actions may be warranted. For further inquiries, please contact Smiths Medical using the following information: Smit

Recall details

Recall number
Z-1681-2025
Status
Open, Classified
Initiated by the company
Apr 10, 2025
Posted by the FDA
May 5, 2025
Recalling firm
Smiths Medical ASD, Inc.
Quantity
313,676 total units
Product code
MEA
FDA submissions
K170982
Distribution
Worldwide - US Nationwide distribution.

Product description

CADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-51; 2) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-51; 3) PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA, Model Number: 21-2112-0100-51; 4) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2112-0300-01; 5) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2111-0300-01; 6) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-51; 7) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 8) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-51; 9) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 10) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, CE ENGLISH 1/EA, Model Number: 21-2111-0400-51; 11) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2111-0300-50; 12) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 13) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 14) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-78; 15) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 16) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2112-0300-00; 17) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-78; 18) PUMP, DEMO, CADD-SOLIS, MODEL 2100, NOT FOR HUMAN USE 1/EA, Model Number: 21-2101-249; 19) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-78; 20) PUMP KIT, CADD-SOLIS, YELLOW, MDL 2100 1/EA, Model Number: 21-2102-51; 21) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, NFHU, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-249; 22) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA, Model Number: 21-2111-0300-231; 23) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-78; 24) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0400-51; 25) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2112-0300-50; 26) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 27) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 28) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 29) PUMP, CADD-SOLIS, MODEL 2110, CE ENGLISH OUS 1/EA, Model Number: 21-2111-0100-50; 30) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 31) PUMP KIT, CADD-SOLIS HPCA, YELLOW KEYPAD, MODEL 2110, CE ENGLISH, OUS 1/EA, Model Number: 21-2112-0100-50; 32) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 33) PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA, Model Number: 21-2112-0400-01.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

All serial numbers. 1 ) 21-2111-0402-51 , UDI-DI: 15019517154290 ; 2 ) 21-2112-0402-51 , UDI-DI: 15019517154993 ; 3 ) 21-2112-0100-51 , UDI-DI: 10610586039232 ; 7 ) 21-2111-0100-51 , UDI-DI: 30610586039205 ; 9 ) 21-2101-51 , UDI-DI: 10610586031854 ; 13 ) 21-2111-0300-00 , UDI-DI: 10610586038792 ; 14 ) 21-2111-0402-78 , UDI-DI: 15019517154887 ; 15 ) 21-2111-0100-00 , UDI-DI: 10610586039195 ; 16 ) 21-2112-0300-00 , UDI-DI: 10610586038822 ; 17 ) 21-2112-0402-78 , UDI-DI: 15019517155013 ; 18 ) 21-2101-249 , UDI-DI: 10610586034787 ; 20 ) 21-2102-51 , UDI-DI: 10610586033933 ; 22 ) 21-2111-0300-231 , UDI-DI: 10610586038785 ; 23 ) 21-2112-0401-78 , UDI-DI: 15019517101034 ; 24 ) 21-2112-0400-51 , UDI-DI: 10610586041051 ; 25 ) 21-2112-0300-50 , UDI-DI: 10610586038860 ; 26 ) 21-2111-0300-00 , UDI-DI: 10610586038792 ; 27 ) 21-2111-0100-00 , UDI-DI: 10610586039195 ; 28 ) 21-2111-0100-51 , UDI-DI: 30610586039205 ; 29 ) 21-2111-0100-50 , UDI-DI: 10610586039256 ; 30 ) 21-2101-51 , UDI-DI: 10610586031854 ; 31 ) 21-2112-0100-50 , UDI-DI: 10610586039270 ; 32 ) 21-2101-00 , UDI-DI: 35019517154904; 33 ) 21-2112-0400-01 , UDI-DI: 10610586041822 ;

Devices named in this recall

8 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

12 devices cleared under K170982 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 96624).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page