CADD recallZ-1681-2025 · Smiths Medical ASD, Inc. · initiated Apr 10, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION dated 4/10/25 was sent to customers. Actions for Users: Inform all affected CADD-Solis and CADD-Solis VIP users of this notice. Provide the instructions below: 1. Ensure all users or potential users of these products are immediately made aware of this notification. 2. Upon receipt of this notification and ongoing during regular maintenance, visually inspect the external condition of the battery pack or Wireless Communication Module and look for evidence of any physical damage. Handle a damaged device with caution. As stated in the battery pack Instructions for Use, if the battery pack housing is cracked or otherwise damaged, replace the battery pack. NEVER use a battery pack that appears damaged. A rechargeable battery pack must be replaced with either another CADD-Solis rechargeable battery pack or with 4 AA batteries. 3. Visually inspect the condition of the battery compartment for evidence of any physical damage. Contact Technical Support for repairs. 4. Always check the battery compartment for fluid or debris before inserting the batteries, and do not allow any fluid or debris to fall into the battery compartment. 5. Complete the electronic Response Form via the link provided within this email notification within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product. 6. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward your customer list to Sedgwick at [email protected]. Sedgwick will notify customers with email addresses via email and will send physical notifications to customers without email addresses in order for them to complete the Response Form Follow up Actions by Smiths Medical: Smiths Medical is continuing to investigate this matter to determine if additional actions may be warranted. For further inquiries, please contact Smiths Medical using the following information: Smit
Recall details
- Recall number
- Z-1681-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 10, 2025
- Posted by the FDA
- May 5, 2025
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- 313,676 total units
- Product code
- MEA
- FDA submissions
- K170982
- Distribution
- Worldwide - US Nationwide distribution.
Product description
CADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-51; 2) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-51; 3) PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA, Model Number: 21-2112-0100-51; 4) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2112-0300-01; 5) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2111-0300-01; 6) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-51; 7) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 8) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-51; 9) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 10) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, CE ENGLISH 1/EA, Model Number: 21-2111-0400-51; 11) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2111-0300-50; 12) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 13) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 14) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-78; 15) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 16) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2112-0300-00; 17) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-78; 18) PUMP, DEMO, CADD-SOLIS, MODEL 2100, NOT FOR HUMAN USE 1/EA, Model Number: 21-2101-249; 19) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-78; 20) PUMP KIT, CADD-SOLIS, YELLOW, MDL 2100 1/EA, Model Number: 21-2102-51; 21) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, NFHU, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-249; 22) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA, Model Number: 21-2111-0300-231; 23) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-78; 24) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0400-51; 25) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2112-0300-50; 26) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 27) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 28) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 29) PUMP, CADD-SOLIS, MODEL 2110, CE ENGLISH OUS 1/EA, Model Number: 21-2111-0100-50; 30) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 31) PUMP KIT, CADD-SOLIS HPCA, YELLOW KEYPAD, MODEL 2110, CE ENGLISH, OUS 1/EA, Model Number: 21-2112-0100-50; 32) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 33) PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA, Model Number: 21-2112-0400-01.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
All serial numbers. 1 ) 21-2111-0402-51 , UDI-DI: 15019517154290 ; 2 ) 21-2112-0402-51 , UDI-DI: 15019517154993 ; 3 ) 21-2112-0100-51 , UDI-DI: 10610586039232 ; 7 ) 21-2111-0100-51 , UDI-DI: 30610586039205 ; 9 ) 21-2101-51 , UDI-DI: 10610586031854 ; 13 ) 21-2111-0300-00 , UDI-DI: 10610586038792 ; 14 ) 21-2111-0402-78 , UDI-DI: 15019517154887 ; 15 ) 21-2111-0100-00 , UDI-DI: 10610586039195 ; 16 ) 21-2112-0300-00 , UDI-DI: 10610586038822 ; 17 ) 21-2112-0402-78 , UDI-DI: 15019517155013 ; 18 ) 21-2101-249 , UDI-DI: 10610586034787 ; 20 ) 21-2102-51 , UDI-DI: 10610586033933 ; 22 ) 21-2111-0300-231 , UDI-DI: 10610586038785 ; 23 ) 21-2112-0401-78 , UDI-DI: 15019517101034 ; 24 ) 21-2112-0400-51 , UDI-DI: 10610586041051 ; 25 ) 21-2112-0300-50 , UDI-DI: 10610586038860 ; 26 ) 21-2111-0300-00 , UDI-DI: 10610586038792 ; 27 ) 21-2111-0100-00 , UDI-DI: 10610586039195 ; 28 ) 21-2111-0100-51 , UDI-DI: 30610586039205 ; 29 ) 21-2111-0100-50 , UDI-DI: 10610586039256 ; 30 ) 21-2101-51 , UDI-DI: 10610586031854 ; 31 ) 21-2112-0100-50 , UDI-DI: 10610586039270 ; 32 ) 21-2101-00 , UDI-DI: 35019517154904; 33 ) 21-2112-0400-01 , UDI-DI: 10610586041822 ;
Devices named in this recall
8 devices in the FDA device registry whose identifier appears in the recall's code information.
- CADD 21-2111-0402-51Icu Medical, Inc. · Pump, infusion, pca4 recalls1,535 reports
- CADD 21-2112-0402-51Icu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery3 recalls469 reports
- CADD 21-2112-0300-00Icu Medical, Inc. · Pump, infusion, pca4 recalls6 reports
- CADD 21-2111-0300-231Icu Medical, Inc. · Pump, infusion, pca4 recalls0 reports
- CADD 21-2112-0100-51Icu Medical, Inc. · Pump, infusion, pca2 recalls59 reports
- CADD 21-2111-0402-78Icu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery3 recalls5 reports
- CADD 21-2112-0402-78Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION3 recalls0 reports
- CADD 2110Icu Medical, Inc. · Pump, infusion, pca2 recalls1 report
Product lines
Other devices under the same FDA submission
12 devices cleared under K170982 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CADD 21-2111-0401-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION4 recalls461 reports
- CADD 21-2112-0402-51LIcu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery3 recalls129 reports
- CADD 21-2111-0402-51LIcu Medical, Inc. · Pump, infusion, pca3 recalls127 reports
- CADD 2110Icu Medical, Inc. · Pump, infusion, pca4 recalls4 reports
- CADD 2110Icu Medical, Inc. · Pump, infusion, pca4 recalls0 reports
- CADD 21-2112-0401-51Icu Medical, Inc. · Pump, infusion, pca1 recall97 reports
- CADD 2110Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall2 reports
- CADD 21-2111-0402-249Icu Medical, Inc. · Pump, infusion, pca1 recall0 reports
- CADD 21-2112-0402-249Icu Medical, Inc. · Set, administration, intravascular1 recall0 reports
- CADD 21-2131-25Icu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery33 reports
- CADD 21-2131-01Icu Medical, Inc. · Set, administration, intravascular21 reports
- CADD 21-2130-51Icu Medical, Inc. · Pump, infusion, pca4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96624).
- Z-1682-2025 · Open, ClassifiedCADD Solis VIP Ambulatory Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE, 1/EA, Model Number: 21-2127-0105-01; 2) PUMP KIT, CADD-SOLIS VIP…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page