Skip to content

Bivona recallZ-2595-2024 · Smiths Medical ASD Inc. · initiated May 29, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

FDA root cause category: Under Investigation by firm

What the company is doing

Smiths Medical issued an URGENT MEDICAL DEVICE NOTIFICATION to its consignees on 05/29/2024 via letter. The notice explained the issue with the device, potential risk and requested the following: "Customer Required Actions: When using the device, all instructions, including warning and cautions contained in the Instructions for Use Documentation must be followed with heightened awareness. Please complete the following actions listed below: 1. Check all inventory locations within your institution for the affected catalog numbers listed in the notification and discontinue use. Discard all affected products following your institution s process for discarding. If discarding is not immediately possible at your facility, then the product should be quarantined until disposal. 2. Share this notification with all potential users of the devices to ensure they are aware of this notification and proposed mitigations. If the devices are used at another location, please ensure this communication is delivered there. 3. Complete and return the attached Customer Response Form to [email protected] within 10 days of receipt to acknowledge your understanding of this notification. Please contact your local representative for assistance with replacement product and/or credit. 3. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]" Questions: Global Complaint Management - 1-(866)-216-8806 or US: [email protected] Customer Service: [email protected] or 1-(800)-258-5361

Recall details

Recall number
Z-2595-2024
Status
Open, Classified
Initiated by the company
May 29, 2024
Posted by the FDA
Aug 23, 2024
Recalling firm
Smiths Medical ASD Inc.
Product code
BTO
FDA submissions
K912469
Distribution
Worldwide distribution.

Product description

PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM UNCUFFED , Product Code/List Number/Item Code 60P025; b) PAED. TRACHEOSTOMY TUBE 3.0MM UNCUFFED , Product Code/List Number/Item Code 60P030; c) PAED. TRACHEOSTOMY TUBE 3.5MM UNCUFFED , Product Code/List Number/Item Code 60P035; d) PAED. TRACHEOSTOMY TUBE 4.0MM UNCUFFED , Product Code/List Number/Item Code 60P040; e) PAED. TRACHEOSTOMY TUBE 4.5MM UNCUFFED , Product Code/List Number/Item Code 60P045; f) PAED. TRACHEOSTOMY TUBE 5.0MM UNCUFFED , Product Code/List Number/Item Code 60P050; g) PAED. TRACHEOSTOMY TUBE 5.5MM UNCUFFED , Product Code/List Number/Item Code 60P055

Code information (models, lots, serial numbers)

Show the full code information (6k characters)

a) UDI/DI 15021312005721, Product Code/List Number/Item Code 60P025, Lot Numbers: 3880717, 3901307, 3939666, 3955786, 3991203, 4015575, 4035758, 4083928, 4110465, 4156650, 4176238, 4209581, 4226268, 4254269, 4331631, 4335055, 4345041, 4348782, 4374468, 4374469, 4389923; b) UDI/DI 15021312005738, Product Code/List Number/Item Code 60P030, Lot Numbers: 3853478, 3868293, 3868294, 3884249, 3897861, 3904660, 3926961, 3936578, 3952364, 3952365, 3952366, 3959293, 3982699, 3997821, 4001215, 4011384, 4015576, 4024940, 4028385, 4040757, 4086004, 4092627, 4136145, 4145692, 4159935, 4185432, 4201393, 4206835, 4216332, 4216333, 4217450, 4223738, 4254270, 4273438, 4276478, 4286815, 4331632, 4335056, 4335057, 4374470, 4374471, 4382705, 4388203, 4389924, 4399210; c) UDI/DI 15021312005745, Product Code/List Number/Item Code 60P035, Lot Numbers: 3840043, 3853480, 3853481, 3864261, 3868247, 3871069, 3877659, 3880688, 3884267, 3891529, 3901310, 3904656, 3911424, 3920956, 3930014, 3939675, 3942522, 3952388, 3952389, 3952390, 3952391, 3959310, 3962581, 3982700, 3982701, 3988152, 3997822, 4001216, 4005677, 4008229, 4008230, 4011385, 4015577, 4026684, 4034945, 4034948, 4077086, 4083939, 4085993, 4095918, 4105252, 4110466, 4116092, 4128015, 4134012, 4134013, 4136146, 4139018, 4147997, 4151027, 4156605, 4163852, 4176208, 4179368, 4196294, 4201394, 4206836, 4209582, 4212790, 4218715, 4221356, 4223739, 4231518, 4237395, 4240163, 4247295, 4254273, 4260076, 4273439, 4276479, 4283252, 4283253, 4286816, 4292179, 4322686, 4327946, 4331633, 4331634, 4335058, 4335059, 4335060, 4335061, 4352799, 4369612, 4369613, 4369614, 4369615, 4374472, 4374473, 4380056, 4381033, 4382712, 4383654, 4388204, 4388206, 4389925, 4393424, 4395802; d) UDI/DI 15021312005752, Product Code/List Number/Item Code 60P040, Lot Numbers: 3840052, 3853482, 3853483, 3861126, 3868273, 3868274, 3874304, 3880711, 3884254, 3887756, 3894882, 3894883, 3904676, 3904677, 3911433, 3911434, 3923724, 3933477, 3939646, 3942486, 3946253, 3949021, 3952360, 3952361, 3952362, 3962569, 3962570, 3962571, 3962572, 3982702, 3988153, 3988154, 3991204, 4001217, 4005678, 4011386, 4015578, 4015580, 4029514, 4029515, 4035761, 4035762, 4035763, 4051642, 4051643, 4053701, 4077099, 4077100, 4082681, 4086016, 4086017, 4095923, 4107611, 4110467, 4113113, 4119089, 4122444, 4128026, 4130997, 4138995, 4142426, 4148005, 4153577, 4156635, 4161984, 4168172, 4179369, 4180501, 4195806, 4196293, 4201395, 4206837, 4206838, 4209583, 4217451, 4218716, 4218717, 4221357, 4226269, 4233979, 4237396, 4240164, 4254275, 4254277, 4273440, 4276480, 4279347, 4283254, 4283255, 4286817, 4289497, 4292180, 4322687, 4327944, 4335062, 4335063, 4335064, 4335065, 4335066, 4348797, 4351249, 4351250, 4369616, 4369617, 4374474, 4374475, 4381034, 4382706, 4382707, 4382708, 4382713, 4382715, 4383655, 4388205, 4388207, 4389926, 4393425; e) UDI/DI 15021312005769, Product Code/List Number/Item Code 60P045, Lot Numbers: 3858169, 3861127, 3868289, 3868290, 3868291, 3884247, 3894909, 3897860, 3904658, 3907984, 3911428, 3911429, 3930056, 3942490, 3946246, 3949024, 3952377, 3952378, 3952379, 3952380, 3955769, 3982703, 3982704, 3982705, 3988155, 3988156, 3988157, 4001218, 4005679, 4008231, 4015582, 4018526, 4018527, 4023793, 4026673, 4029516, 4035764, 4051639, 4051640, 4053716, 4060205, 4071970, 4077069, 4077070, 4083907, 4086002, 4086003, 4092626, 4096837, 4107598, 4110468, 4119117, 4125816, 4130978, 4138991, 4142422, 4145691, 4151044, 4153613, 4159933, 4161970, 4168139, 4176217, 4180500, 4185433, 4195807, 4196295, 4197626, 4201396, 4206839, 4206840, 4209584, 4212791, 4216334, 4218718, 4218719, 4221358, 4226270, 4231519, 4237397, 4240165, 4240166, 4254278, 4260077, 4260078, 4273441, 4276481, 4283256, 4286818, 4289498, 4292181, 4322688, 4331635, 4331636, 4335067, 4335068, 4335069, 4352800, 4352801, 4352802, 4357224, 4357226, 4374476, 4374477, 4374478, 4381035, 4381036, 4382714, 4383656, 4388208, 4388209, 4389927; f) UDI/DI 15021312005776, Product Code/List Number/Item Code 60P050, Lot Numbers: 3853484, 3853485, 3868201, 3868202, 3871066, 3884260, 3891527, 3894890, 3904663, 3904664, 3907966, 3918021, 3918022, 3926969, 3942507, 3946271, 3946272, 3952401, 3952402, 3952403, 3955781, 3982706, 3982707, 3982708, 3988158, 3988159, 3988160, 3988161, 3991205, 3997823, 4005680, 4005681, 4005682, 4009521, 4015583, 4015586, 4015587, 4027994, 4027995, 4040758, 4040759, 4077068, 4080313, 4085997, 4085998, 4092634, 4097839, 4107583, 4107584, 4113118, 4122446, 4125824, 4136147, 4139000, 4147981, 4151023, 4159937, 4168184, 4173735, 4179370, 4196296, 4204825, 4206843, 4209585, 4209586, 4216335, 4218720, 4218721, 4221359, 4226271, 4233980, 4254279, 4254280, 4273442, 4276482, 4279348, 4283257, 4283258, 4286819, 4289496, 4292182, 4322689, 4327947, 4331637, 4335070, 4335071, 4335072, 4335073, 4345042, 4351251, 4352804, 4369620, 4369621, 4369622, 4374479, 4374480, 4380057, 4382709, 4383657, 4388210, 4388211, 4389928, 4393426, 4395804, 4399218, 4400699, 4400700; g) UDI/DI 15021312005783, Product Code/List Number/Item Code 60P055, Lot Numbers: 3853487, 3853488, 3868223, 3880715, 3897864, 3908004, 3908005, 3923725, 3930031, 3936576, 3939631, 3942495, 3949031, 3952372, 3952373, 3952374, 3955790, 3982709, 3982710, 3988162, 3988163, 3991206, 3991207, 4001219, 4005683, 4005684, 4015588, 4015590, 4015591, 4015592, 4048027, 4051649, 4053708, 4066296, 4071976, 4077092, 4080784, 4083916, 4091395, 4105255, 4107591, 4113096, 4122455, 4130992, 4142434, 4145699, 4153620, 4159968, 4168153, 4173736, 4180465, 4185434, 4201397, 4204826, 4206844, 4209587, 4212792, 4216336, 4218722, 4223740, 4233981, 4254281, 4254282, 4254283, 4273443, 4276483, 4279349, 4283259, 4286820, 4289499, 4292183, 4322690, 4327941, 4327942, 4335075, 4335076, 4335077, 4351252, 4352803, 4374481, 4374482, 4374483, 4380058, 4380059, 4380060, 4381037, 4382716, 4383658, 4388212, 4388213, 4389929, 4393427, 4399217, 4399219

Devices named in this recall

7 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

19 devices cleared under K912469 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 94852).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page