Cytomics FC500Recalls & reported problems across all models
Beckman Coulter, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1,337
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls
- 0
We found 2 models of Cytomics FC500 by Beckman Coulter, Inc. in the FDA's device registry (GUDID). Together they are named in 1,337 adverse event reports in the FDA's MAUDE database, received between Sep 5, 2018 and Nov 20, 2024 — classified as 4 injuries, 1,333 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1,330 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Cytomics FC500 problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1,330
- Device Remains Activated1
- Electrical /Electronic Property Problem1
- Fire1
- Inadequate or Insufficient Training1
- No Apparent Adverse Event1
- Output Problem1
- Retraction Problem1
Patient problems
- No Consequences Or Impact To Patient1,125
- No Clinical Signs, Symptoms or Conditions210
- Head Injury1
- Test Result1
Cytomics FC500 models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Cytomics FC500 UPS (CXP)Counter, Differential Cell1,196 reports
- Cytomics FC500 UPS (MPL)Counter, Differential Cell141 reports
Recent Cytomics FC500 adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED ABNORMAL HISTOGRAMS ON THE FC500 FLOW CYTOMETER. THE FIELD SERVICE ENGINEER (FSE) FOUND THERE WAS A TARPON AMP BOARD ERROR AT FORWARD SCATTER (FS). THE EVENT OCCURRED DURING START THE FLOW CHECK (QUALITY CONTROL) RUN. THERE WAS NO REPORT OF ERRONEOUS PATIENT RESULTS ASSOCIATED WITH THE EVENT. THERE WAS NO REPORTED DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20738111 ↗MalfunctionJul 18, 2024No Apparent Adverse Event; No Apparent Adverse Event
THE CUSTOMER REPORTED AMP BOARD ERRORS AT THE FL5 FLUORESCENCE CHANNEL ON THEIR FC500 FLOW CYTOMETER WHILE RUNNING FLOWCHECK AT START-UP. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT. PATIENT SAMPLES WERE NOT RUN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19778802 ↗MalfunctionJul 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED AMP BOARD ERRORS AT THE FL5 FLUORESCENCE CHANNEL ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19776768 ↗MalfunctionMar 20, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED FL3 AND FL5 AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT. THERE WAS NO REPORT OF ERRONEOUS RESULTS GENERATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18944227 ↗MalfunctionJul 20, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED TARPON AMP BOARD ERRORS ON THE "FORWARD" SCATTER (FS), FL1 AND A FLOW CHECK SIGNAL DRIFT AT FL3 ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17364165 ↗MalfunctionJul 12, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED NO SIGNAL AT THE FL4 POSITION ON THEIR FC 500 FLOW CYTOMETER WHILE RUNNING FLOW CHECK. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17311809 ↗MalfunctionJun 26, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED FL4 PE TARPON AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17203053 ↗MalfunctionJun 8, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED INTERMITTENT SIGNAL LOSS FOR FL4 AND UNSTABLE SIGNAL FOR FL3 WHEN RUNNING FLOWCHECK ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17091997 ↗MalfunctionMay 8, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED FL1 AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER WHILE RUNNING FLOW CHECK. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16887916 ↗MalfunctionApr 28, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED AMP BOARD ERRORS AT THE FOLLOWING FLUORESCENCE CHANNELS (FL2, FL3 AND FL4) POSITIONS ON THEIR FC500, WHILE RUNNING FLOW CHECK AND QUALITY CONTROL. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16833605 ↗MalfunctionApr 24, 2023Fire; Fire
THE CUSTOMER REPORTED HEARING A POPPING ELECTRICAL NOISE ON THEIR FC 500 FLOW CYTOMETER. THE UNIT (POWER SUPPLY CART) CAUGHT FIRE AS IT SAT IDLE. THE CUSTOMER REPORTED SEEING FLAMES THAT WERE PUT OUT WITH A FIRE EXTINGUISHER. THERE WAS NO DIRECT HARM TO LABORATORY OR LABORATORY PERSONNEL. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16799586 ↗MalfunctionApr 10, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED FL4 TARPON AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER, WHILE PROCESSING BOTH FLOW CHECK AND QUALITY CONTROL. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16709296 ↗MalfunctionMar 29, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED INTERMITTENT FL4 X-MEAN SIGNAL SHIFT AND FL4 TARPON AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER, WHILE PROCESSING BOTH FLOW CHECK AND QUALITY CONTROL. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16638285 ↗MalfunctionMar 9, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED TARPON AMP BOARD ERRORS AT THE FL3 AND SS (SIDE SCATTER) POSITION ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16512059 ↗MalfunctionJan 4, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED AMP BOARD ERROR AT THE FLUORESCENCE CHANNELS (FL1,FL2,FL3, FL4 AND FL5) WHEN PROCESSING FLOW CHECK, QUALITY CONTROL. AND PATIENT SAMPLES ON THEIR FC500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16099811 ↗
Other Beckman Coulter, Inc. products
- DxC AU4 models · 632 reports · 8 recalls
- MicroScan10 models · 7 reports · 9 recalls
- Synchron7 models · 4 reports · 5 recalls
- DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module2 models · 5 reports · 4 recalls
- UniCel DxH4 models · 74 reports · 2 recalls
- Access AccuTnI+32 models · 165 reports
- REMISOL Advance2 models · 0 reports · 2 recalls
- Access TSH (3rd IS)2 models · 7 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page