Skip to content

UniCel DxH 690T Coulter Cellular Analysis System C34520Recalls & reported problems

Beckman Coulter, Inc. · DxH 690T Hematology Analyzer, DxH 690T · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15099590729110), received between Jul 15, 2021 and Jul 15, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 13, 2025Class IIOpen, Classified

    Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

    Z-2573-2025 · Root cause: Component change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2021Leak/Splash; Leak/Splash

    THE CUSTOMER REPORTED THAT THE POWER STRIP WHERE THE DXH690T HEMATOLOGY INSTRUMENT WAS PLUGGED OVERLOADED AND GENERATED SPARKS. THE FIRE DEPARTMENT WAS CALLED AND THE LABORATORY EVACUATED. THERE WAS NO INDICATION OF ERRONEOUS RESULTS BEING REPORTED OUTSIDE THE LABORATORY. NO REPORTED CHANGE OR EFFECT ON PATIENT TREATMENT. THERE WAS NEITHER DEATH OR SERIOUS INJURY ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12172832 ↗

Other Counter, Differential Cell devices

All Counter, Differential Cell devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page