Portico Solo™ PRTSOLO-19 (UDI 15415067028157)Recalls & reported problems

St. Jude Medical, Inc. · Re-Collapsible Introducer Hydrophillic Coated 11F Comparable Sheath ID 19F Expanded Profile · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15415067028157), received between May 3, 2017 and Dec 14, 2018 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult to Remove2
  • Deflation Problem1

Patient problems

  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 14, 2018Deflation Problem; Deflation Problem

    THE USER FACILITY REPORTED A 11F PORTICO SOLOPATH WAS USED FOR A PORTICO IMPLANT. BUT THEY WERE UNABLE TO RE-COLLAPSE THE PORTICO SOLO FOR REMOVAL AFTER THE SUCCESSFULLY IMPLANTATION OF A 25MM PORTICO VALVE. SO THE PHYSICIAN REMOVED IT DIRECTLY. THE PATIENT WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8166269 ↗
  • MalfunctionJun 15, 2017Difficult to Remove; Difficult to Remove

    THE USER FACILITY REPORTED THAT THERE WAS A METAL LOOP VISIBLE ON THE SHEATH. THE FOLLOWING INFORMATION WAS REPORTED: A PORTICOSOLO-19 DEVICE WAS USED DURING THE TAVI PROCEDURE; AFTER REMOVING THE PORTICO SOLO DEVICE, THE USER NOTICED THAT A METAL LOOP WAS VISIBLE OUTSIDE OF THE LOCK AT THE END OF THE SHEATH; THE PATIENT WAS REPORTED TO BE STABLE; AND NO INJURIES WERE OBSERVED ON THE PATIENT'S VESSELS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6642950 ↗
  • InjuryMay 3, 2017Difficult to Remove; Difficult to Remove

    THE USER FACILITY REPORTED THAT THERE WAS EXTENSIVE UNFOLDING OF THE SHEATH. FOLLOW UP COMMUNICATION WITH THE USER FACILITY CONFIRMED THE FOLLOWING INFORMATION: A PROGLIDE WAS CUT OUT OF THE PRE-CLOSURE SITE; DUE TO INTERNAL BALLOONING A STENT GRAFT IMPLANTATION WAS NOT NECESSARY; AND THE PATIENT IS STABLE.

    Patient problems: No Consequences Or Impact To Patient