3Dimensions recallZ-2168-2026 · Hologic, Inc · initiated Apr 8, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Hologic, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
FDA root cause category: Device Design
What the company is doing
On April 8, 2026 URGENT FIELD SAFETY NOTICE letters were sent to customers. Customers were informed of implementing an updated recurring maintenance inspection for detecting loosened, missing, or broken bolts within the VTA. Hologic Service will reach out to schedule your service appointment and perform the required inspection. Following the initial inspection, recurring inspections are required to be conducted at least every 6 months to ensure continued safe/proper operation of the system. Future inspections will consist of torque and visual confirmation that all bolts are present and that there is no indication of loosening. Actions: You may continue to use your 3Dimensions or Selenia Dimensions for its intended use. If at any time during system operation, you observe unexpected C-Arm movement, please discontinue use and contact Hologic s Technical Support immediately. Report suspected device complaints and/or adverse events to Hologic and the FDA (contact details below). After initial inspection by Hologic, it is required to have the inspection performed at least every 6 months as well as the other defined maintenance and service activities which are critical for continued safe and effective operation of the Selenia Dimensions or 3Dimensions system. Forward this notice to anyone in your facility that needs to be informed. Post a copy of this notice in a visible area near the affected devices where operators can view for awareness and please keep a copy for your records.
Recall details
- Recall number
- Z-2168-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 8, 2026
- Posted by the FDA
- May 12, 2026
- Recalling firm
- Hologic, Inc
- Quantity
- 14837 units
- Product code
- OTE
- FDA submissions
- P080003
- Distribution
- US Nationwide distribution.
Product description
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-INTL3D, 3DM-SYS-INTL3D-MOB, 3DM-SYS-INTL3D-NS, 3DM-SYS-STD, 3DM-SYS-STD-MOB, 3DM-SYS-STD-NS.
Code information (models, lots, serial numbers)
"[Model No.; UDI]: [3DM-SYS-INTL2D;15420045510609], [3DM-SYS-INTL2D-MOB;15420045510593], [3DM-SYS-INTL2D-NS;15420045510586], [3DM-SYS-INTL3D;15420045510579], [3DM-SYS-INTL3D-MOB;15420045510562], [3DM-SYS-INTL3D-NS;15420045510555], [3DM-SYS-STD;15420045510524], [3DM-SYS-STD-MOB;15420045510531], [3DM-SYS-STD-NS;15420045510548]; All Serial No.
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- 3Dimensions 3DM-SYS-STDHologic, Inc. · Digital breast tomosynthesis1 recall57 reports
- 3Dimensions 3DM-SYS-INTL3DHologic, Inc. · Digital breast tomosynthesis1 recall5 reports
- 3Dimensions 3DM-SYS-INTL2D-MOBHologic, Inc. · Digital breast tomosynthesis1 recall5 reports
- 3Dimensions 3DM-SYS-STD-NSHologic, Inc. · Digital breast tomosynthesis1 recall0 reports
- 3Dimensions 3DM-SYS-INTL3D-NSHologic, Inc. · Digital breast tomosynthesis1 recall0 reports
- 3Dimensions 3DM-SYS-INTL3D-MOBHologic, Inc. · Digital breast tomosynthesis1 recall0 reports
- 3Dimensions 3DM-SYS-INTL2D-NSHologic, Inc. · Digital breast tomosynthesis1 recall0 reports
Product lines
Other devices under the same FDA submission
8 devices cleared under P080003 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Selenia Dimensions SDM-SYS-6000-3DHologic, Inc. · Digital breast tomosynthesis1 recall29 reports
- Selenia Dimensions SDM-SYS-9000-3DHologic, Inc. · Digital breast tomosynthesis1 recall16 reports
- Selenia Dimensions SDM-SYS-6000-2DHologic, Inc. · Digital breast tomosynthesis1 recall5 reports
- Selenia Dimensions SDA-SYS-3000-2DHologic, Inc. · Digital breast tomosynthesis1 recall4 reports
- Selenia Dimensions SDM-SYS-9000-2DHologic, Inc. · Digital breast tomosynthesis1 recall2 reports
- 3Dimensions ASY-10559Hologic, Inc. · Digital breast tomosynthesis1 recall0 reports
- 3Dimensions ASY-11748-CHDHologic, Inc. · Digital breast tomosynthesis1 recall0 reports
- Selenia Dimensions 3D System 3D softwareHologic, Inc. · Digital breast tomosynthesis14 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98566).
- Z-2167-2026 · Open, ClassifiedSELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page