VIEW PLATING SYSTEM, REDUCE FRACTURE PLATING SYSTEM 1 (REF 002-D-27014)Recalls & reported problems
Orthofix Srl · NON-LOCKING SCREW, D 2.7MM X 14MM · Class 2
All models of VIEW PLATING SYSTEM, REDUCE FRACTURE PLATING SYSTEM (140)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 18056099645812), received between Jan 23, 2018 and Jan 23, 2018 — classified as 2 injuries.
The most frequently reported specific device problem is “Mechanical Problem” (2 reports).
1 FDA recall record name this device's identifier.
14 other recalls cover devices cleared under the same FDA submission (K140876); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Apr 6, 2018Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Z-2121-2018 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem2
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 23, 2018Mechanical Problem; Mechanical Problem
THE INFORMATION PROVIDED BY THE LOCAL DISTRIBUTOR INDICATES: - HOSPITAL NAME: (B)(6); - SURGEON NAME: (B)(6); - DATE OF SURGERY: (B)(6) 2017; - BODY PART TO WHICH DEVICE WAS APPLIED: FOOT; - SURGERY DESCRIPTION: NO RESPONSE; - PATIENT INFORMATION: (B)(6), MALE; - PROBLEM OBSERVED DURING: CLINICAL USE ON PATIENT; - TYPE OF PROBLEM: DEVICE FUNCTIONAL PROBLEM. - EVENT DESCRIPTION: SCREWS FAILED DURING SURGERY, (B)(4) THREADS STRIPPED DURING INSERTION BUT WAS ABLE TO REMOVE (PLEASE KINDLY REFER TO MFR REPORT 9680825-2018-0002). (B)(4), ONE THREADS STRIPPED DURING INSERTION AND THE OTHER HEAD BROKE OFF AT THE NECK DURING INSERTION (PLEASE KINDLY REFER TO MFR REPORTS 9680825-2018-00003 AND 9680825-2018-00004). THEY REMAIN IN PATIENT. THE COMPLAINT REPORT FORM ALSO INDICATES: - THE DEVICE FAILURE DID NOT HAVE ANY ADVERSE EFFECTS ON PATIENT; - THE SURGERY WAS NOT COMPLETED WITH THE DEVICE; - A REPLACEMENT DEVICE WAS IMMEDIATELY AVAILABLE TO COMPLETE THE SURGERY; - THE EVENT DID NOT LEAD TO A CLINICALLY RELEVANT INCREASE IN THE DURATION OF THE SURGICAL PROCEDURE (10 MINUTES); - AN ADDITIONAL SURGERY WAS NOT REQUIRED; - COPIES OF THE OPERATIVE REPORT ARE AVAILABLE (NOT RECEIVED); - COPIES OF
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7211655 ↗InjuryJan 23, 2018Mechanical Problem; Mechanical Problem
THE INFORMATION PROVIDED BY THE LOCAL DISTRIBUTOR INDICATES: HOSPITAL NAME: (B)(6); SURGEON NAME: (B)(6); DATE OF SURGERY: (B)(6) 2017; BODY PART TO WHICH DEVICE WAS APPLIED: FOOT; SURGERY DESCRIPTION: NO RESPONSE; PATIENT INFORMATION: (B)(6) YEARS, MALE; PROBLEM OBSERVED DURING: CLINICAL USE ON PATIENT; TYPE OF PROBLEM: DEVICE FUNCTIONAL PROBLEM. EVENT DESCRIPTION: SCREWS FAILED DURING SURGERY, 002-E-27018 THREADS STRIPPED DURING INSERTION BUT WAS ABLE TO REMOVE (PLEASE KINDLY REFER TO MFR REPORT 9680825-2018-0002). THE 002-E-27014, ONE THREADS STRIPPED DURING INSERTION AND THE OTHER HEAD BROKE OFF AT THE NECK DURING INSERTION (PLEASE KINDLY REFER TO MFR REPORTS 9680825-2018-00003 AND 9680825-2018-00004). THEY REMAIN IN PATIENT. THE COMPLAINT REPORT FORM ALSO INDICATES: THE DEVICE FAILURE DID NOT HAVE ANY ADVERSE EFFECTS ON PATIENT; THE SURGERY WAS NOT COMPLETED WITH THE DEVICE; A REPLACEMENT DEVICE WAS IMMEDIATELY AVAILABLE TO COMPLETE THE SURGERY; THE EVENT DID NOT LEAD TO A CLINICALLY RELEVANT INCREASE IN THE DURATION OF THE SURGICAL PROCEDURE (10 MINUTES); AN ADDITIONAL SURGERY WAS NOT REQUIRED; COPIES OF THE OPERATIVE REPORT ARE AVAILABLE (NOT RECEIVED); COPIES OF THE X-RAY I
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7211654 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140876. One submission often covers many models, so check the product and lot details in each record.
- Z-2132-2018 · Apr 6, 2018 · TerminatedStraight Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Holes (Part Number): a) 3 Hole (003-D-01001); b) 4 Hole (003-D-02001); c) 6 Hole (003-D-03001)
- Z-2126-2018 · Apr 6, 2018 · TerminatedCC Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-G-01020); b) Medium (002-G-01025); c) Large (002-G-01030)
- Z-2135-2018 · Apr 6, 2018 · TerminatedY-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (003-C-01001); b) Large (003-C-02001)
- Z-2131-2018 · Apr 6, 2018 · TerminatedMTP Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-B-01001); b) Medium (002-B-02002); c) Large (002-B-03003)
- Z-2134-2018 · Apr 6, 2018 · TerminatedT-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (003-B-01001); b) Large (003-B-02001)
- Z-2129-2018 · Apr 6, 2018 · TerminatedLapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM Step/Left (002-C-01000L); b) Small/0MM Step/Right (002-C-0100
- Z-2119-2018 · Apr 6, 2018 · TerminatedNon-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (0002-D-20008); b) D 2.0MM X 10MM (002-D-20010); c) D 2.0MM X 12MM (002-D
- Z-2127-2018 · Apr 6, 2018 · TerminatedEvans Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 6MM Spacer (002-H-00006); b) 8MM Spacer (022-H-00008); c) 10MM Spacer (002-H-00010); d) 12M
- Z-2123-2018 · Apr 6, 2018 · TerminatedNon-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a)D 3.2MM X 10MM (002-D-32010) b) D 3.2MM X 12MM (002-D-32012) c) D 3.2MM X 14MM (002-D-320
- Z-2120-2018 · Apr 6, 2018 · TerminatedLocking Screw, D 2.0MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (002-E-20008); b) D 2.0MM X 10MM (002-E-20010); c) D 2.0MM X 1
- Z-2124-2018 · Apr 6, 2018 · TerminatedLocking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 3.2MM X 10MM (002-E-32010) b) D 3.2MM X 12MM (002-E-32012) c) D 3.2MM X 14MM (002-E-32014)
- Z-2133-2018 · Apr 6, 2018 · TerminatedTN Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-F-01020); b) Medium (002-F-01025); c) Large (002-F-01030)
- Z-2130-2018 · Apr 6, 2018 · TerminatedL-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Left or Right (Part Number): a) Left (003-E-00001L); b) Right (003-E-00001R)
- Z-2122-2018 · Apr 6, 2018 · TerminatedLocking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 2.7MM X 8MM (002-E-27008), b) D 2.7MM X 10MM (002-E-27010); c) D 2.7MM X 12MM (002-E-27012
Other Screw, Fixation, Bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page