Skip to content

VIEW PLATING SYSTEM, REDUCE FRACTURE PLATING SYSTEM 1 (REF 002-D-32020)Recalls & reported problems

Orthofix Srl · NON-LOCKING SCREW, D 3.2MM X 20MM · Class 2

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

14 other recalls cover devices cleared under the same FDA submission (K140876); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Apr 6, 2018Terminated

    Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

    Z-2123-2018 · Root cause: Device Design · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K140876. One submission often covers many models, so check the product and lot details in each record.

  • Z-2121-2018 · Apr 6, 2018 · Terminated
    Non-Locking Screw, D 2.7MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Numbers): a) D 2.7MM X 8MM (002-D-27008); b) D 2.7MM X 10MM (002-D-27010); c) D 2.7
  • Z-2130-2018 · Apr 6, 2018 · Terminated
    L-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Left or Right (Part Number): a) Left (003-E-00001L); b) Right (003-E-00001R)
  • Z-2127-2018 · Apr 6, 2018 · Terminated
    Evans Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 6MM Spacer (002-H-00006); b) 8MM Spacer (022-H-00008); c) 10MM Spacer (002-H-00010); d) 12M
  • Z-2133-2018 · Apr 6, 2018 · Terminated
    TN Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-F-01020); b) Medium (002-F-01025); c) Large (002-F-01030)
  • Z-2134-2018 · Apr 6, 2018 · Terminated
    T-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (003-B-01001); b) Large (003-B-02001)
  • Z-2131-2018 · Apr 6, 2018 · Terminated
    MTP Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-B-01001); b) Medium (002-B-02002); c) Large (002-B-03003)
  • Z-2129-2018 · Apr 6, 2018 · Terminated
    Lapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM Step/Left (002-C-01000L); b) Small/0MM Step/Right (002-C-0100
  • Z-2135-2018 · Apr 6, 2018 · Terminated
    Y-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (003-C-01001); b) Large (003-C-02001)
  • Z-2122-2018 · Apr 6, 2018 · Terminated
    Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 2.7MM X 8MM (002-E-27008), b) D 2.7MM X 10MM (002-E-27010); c) D 2.7MM X 12MM (002-E-27012
  • Z-2126-2018 · Apr 6, 2018 · Terminated
    CC Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-G-01020); b) Medium (002-G-01025); c) Large (002-G-01030)
  • Z-2119-2018 · Apr 6, 2018 · Terminated
    Non-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (0002-D-20008); b) D 2.0MM X 10MM (002-D-20010); c) D 2.0MM X 12MM (002-D
  • Z-2120-2018 · Apr 6, 2018 · Terminated
    Locking Screw, D 2.0MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (002-E-20008); b) D 2.0MM X 10MM (002-E-20010); c) D 2.0MM X 1
  • Z-2132-2018 · Apr 6, 2018 · Terminated
    Straight Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Holes (Part Number): a) 3 Hole (003-D-01001); b) 4 Hole (003-D-02001); c) 6 Hole (003-D-03001)
  • Z-2124-2018 · Apr 6, 2018 · Terminated
    Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 3.2MM X 10MM (002-E-32010) b) D 3.2MM X 12MM (002-E-32012) c) D 3.2MM X 14MM (002-E-32014)

Other Screw, Fixation, Bone devices

All Screw, Fixation, Bone devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.