Skip to content

VIEW PLATING SYSTEM, REDUCE FRACTURE PLATING SYSTEM 1 (REF 003-F-18001)Recalls & reported problems

Orthofix Srl · BOX PLATE, 18MM · Class 2 · Implant

All models of VIEW PLATING SYSTEM, REDUCE FRACTURE PLATING SYSTEM (140)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Apr 6, 2018Class IITerminated

    Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

    Z-2125-2018 · Root cause: Device Design · Recall details · FDA record ↗

Other Plate, Fixation, Bone devices

All Plate, Fixation, Bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page